FDA Enforcement Class II Ongoing

Trevo NXT ProVue 3 x 32, US, Catalog: 90312 Kits: TREVO NXT 3X32+TRAK21 2-PACK - CE, Catalog: 91412 TREVO NXT 4X28+TRAK21 2-PACK - CE, Catalog: 91413 TREVO NXT 4X41+TRAK21 2-PACK - CE, Catalog: 91414 TREVO NXT 6X37+TRAK21 2-PACK - CE, Catalog: 91415 TREVO NXT 3X32+PRO14 2-PACK - CE, Catalog: 91416 TREVO NXT 4X28+TRAK21+CAT6 3-PACK - CE, Catalog: SFP914170 TREVO NXT 6X37+TRAK21+CAT6 3-PACK - CE, Catalog: SFP914190

Recall: Z-1905-2024 · Reported June 5, 2024

Enforcement

Recall Number
Z-1905-2024
Event ID
94437
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Stryker Neurovascular
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 5, 2024
Initiation Date
April 23, 2024
Classification Date
May 28, 2024
Address
47900 Bayside Pkwy, Fremont, CA, 94538-6515, United States

Description

Trevo NXT ProVue 3 x 32, US, Catalog: 90312 Kits: TREVO NXT 3X32+TRAK21 2-PACK - CE, Catalog: 91412 TREVO NXT 4X28+TRAK21 2-PACK - CE, Catalog: 91413 TREVO NXT 4X41+TRAK21 2-PACK - CE, Catalog: 91414 TREVO NXT 6X37+TRAK21 2-PACK - CE, Catalog: 91415 TREVO NXT 3X32+PRO14 2-PACK - CE, Catalog: 91416 TREVO NXT 4X28+TRAK21+CAT6 3-PACK - CE, Catalog: SFP914170 TREVO NXT 6X37+TRAK21+CAT6 3-PACK - CE, Catalog: SFP914190

Reason

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

Code Info

Catalog/UDI-DI/Lot: 90312/07613327299922/118881; 91412/07613327500684/48483279, 48483281, 48483294, 48696618, 48696619, 48696620, 48715713, 48834692, 48834695, 48834696; 91413/07613327500691/48483280, 48483284, 48483289, 48483291, 48483316, 48483317, 48696613, 48696622, 48696623, 48715711, 48715715; 91414/07613327500707/48447776, 48447778, 48447779, 48483290, 48483292, 48483293, 48483315, 48483318, 48483320, 48696624, 48715714, 48834690, 48834691, 48834694, 48834697, 48852726, 48852730, 48852731, 48852732, 48852734, 48852735, 48852737, 48852738; 91415/07613327500714/48483277, 48483283, 48483295, 48483296, 48483311, 48483312, 48696621, 48715712, 48715716, 48834688, 48834693, 48834698; 91416/07613327507904/47328863, 47328865, 47351815, 47351816, 47374814, 48500988, 48501006, 48501007, 48501008, 48501009, 48501010, 48501011, 48501012, 48501013, 48501014, 48540277, 48556780, 48556783, 48558375, 48588482, 48589876, 48641524, 50065509, 50065520, 50193572, 50260581, 50327385, 50327403, 50375186, 50375187, 50422577, 50422588, 50448274, 50738892, 50762691, 50762721, 50762722, 50786687, 50797491, 50797494, 50797495, 50886508, 51320490, 51320491, 51379394, 51379399, 51379400, 51379401, 51379402, 51379403, 51379404, 51379405, 51379406, 51379408, 51467927, 51489221, 51489230, 51489233, 51489234; SFP914170/07613327508055/48834689; SFP914190/07613327508079/48945914

Distribution

Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.

Quantity

204