FDA Enforcement Class II Ongoing

Tibial Tray Nonporous A Size 4; Model # PN 162-1400-A Material: Ti6Al4V.

Recall: Z-1904-2020 · Reported May 13, 2020

Enforcement

Recall Number
Z-1904-2020
Event ID
85157
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ortho Development Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
May 13, 2020
Initiation Date
May 30, 2019
Classification Date
May 5, 2020
Address
12187 S Business Park Dr, N/A, Draper, UT, 84020-8663, United States

Description

Tibial Tray Nonporous A Size 4; Model # PN 162-1400-A Material: Ti6Al4V.

Reason

The locking mechanism of the tibial tray exhibited a manufacturing defect in which a uniform ridge was cut along the inner wall of the "T" and "Y" junctions, which may cause resistance when the Tibial Insert was being assembled.

Code Info

Lot Number: A204324

Distribution

US: CA, FL, OH, PA, TX, and UT. OUS: Japan.

Quantity

29