FDA Enforcement
Class II
Ongoing
Tibial Tray Nonporous A Size 2; Model # PN 162-1200-A; Material: Ti6Al4V.
Recall: Z-1903-2020
·
Reported May 13, 2020
Enforcement
- Recall Number
- Z-1903-2020
- Event ID
- 85157
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Ortho Development Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- May 13, 2020
- Initiation Date
- May 30, 2019
- Classification Date
- May 5, 2020
- Address
- 12187 S Business Park Dr, N/A, Draper, UT, 84020-8663, United States
Description
Tibial Tray Nonporous A Size 2; Model # PN 162-1200-A; Material: Ti6Al4V.
Reason
The locking mechanism of the tibial tray exhibited a manufacturing defect in which a uniform ridge was cut along the inner wall of the "T" and "Y" junctions, which may cause resistance when the Tibial Insert was being assembled.
Code Info
Lot Number: A200590
Distribution
US: CA, FL, OH, PA, TX, and UT. OUS: Japan.
Quantity
27