FDA Enforcement Class I Ongoing

Heartware Medtronic HVAD System for cardiac use. Including the following parts: (a) Controller / Controller Kits, Product numbers: 1400, 1401, 1403, 1407, 1420 (b) DC Adapter, Product numbers: 1435, 1440 (c) AC Adapter, Product numbers: 1425, 1430 (d) Battery Pack, Product number: 1650DE Indication for Use for OUS HVAD System: The HVAD System is intended for use in patients at risk of death from refractory end-stage heart failure. The HVAD System is designed for in-hospital and out-of-hospital settings, including transportation via fixed wing aircraft or helicopter.

Recall: Z-1903-2018 · Reported June 6, 2018

Enforcement

Recall Number
Z-1903-2018
Event ID
80056
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Heartware
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 6, 2018
Initiation Date
May 2, 2018
Classification Date
May 31, 2018
Address
14400 NW 60th Ave, N/A, Miami Lakes, FL, 33014-2807, United States

Description

Heartware Medtronic HVAD System for cardiac use. Including the following parts: (a) Controller / Controller Kits, Product numbers: 1400, 1401, 1403, 1407, 1420 (b) DC Adapter, Product numbers: 1435, 1440 (c) AC Adapter, Product numbers: 1425, 1430 (d) Battery Pack, Product number: 1650DE Indication for Use for OUS HVAD System: The HVAD System is intended for use in patients at risk of death from refractory end-stage heart failure. The HVAD System is designed for in-hospital and out-of-hospital settings, including transportation via fixed wing aircraft or helicopter.

Reason

Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary power source and/or unexpected audible tones (beeping). The interruption typically lasts 1-2 seconds, due to oxidation between power source connector and socket.

Code Info

ALL SERIAL NUMBERS UPN/GTIN: (a) 1400 and 1401: distributed prior to GTN requirement 1403: 00888707000116, 00888707000475 1407: '00888707000291, '00888707000727, '00888707000710, 00888707000734, '00888707000741, '00888707000765, '00888707000758, '00888707000789, '00888707000772, '00888707000499, '00888707000796, '00888707000802, '00888707000819, '00888707000482, '00888707004534, '0888707001670, '00888707001656, '00888707001663, '00888707001670, '00888707001687, '00888707001694, '00888707001700, '00888707001717, '00888707001724, '00888707001731, '00888707001748, '00888707001755, '00888707001762, '00888707002813, '00888707002820, '00888707002837, '00888707002844, '00888707002851, '00888707001472 1420: '00888707000420, '00888707000437, '00888707002660 (b) 1435: '00888707000109 1440: '00888707000260, '00888707001885, '00888707001496, '00888707002745 (c) 1425: '00888707000093 1430: '00888707000307, '00888707000826, '00888707000833, '00888707000840, '00888707000857, '00888707000871, '00888707000864, '00888707000888, '00888707000901, '00888707000505, '00888707000895, '00888707000918, '00888707001489, '00888707001779, '00888707001786, '00888707001793, '00888707001809, '00888707001816, '00888707001823, '00888707001830, '00888707001847, '00888707001854, '00888707001861, '00888707001878, '00888707002738, '00888707002769, '00888707002776, '00888707002783, '00888707002790, '00888707002806, '00888707004817 (d) 1650DE: 00888707000369, 00888707001373, '00888707001588, '00888707001366, '00888707002684 '00888707002646, '00888707002653, '00888707000376

Distribution

Worldwide Distribution including US Nationwide, Argentina, Australia, Austria, Belarus, Belgium, Brazil, Canada, Chile, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Kazakhstan, Korea, , Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, United Arab Emirates, and Vietnam.

Quantity

175878 devices