Delfi Medical Innovations, VariFit Contour Cuffs, single use adult thigh cuff, single port, single bladder, REF 9-9300-003. Rx Only, Sterile
Enforcement
- Recall Number
- Z-1901-2020
- Event ID
- 84741
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Delfi Medical Innovations, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- May 13, 2020
- Initiation Date
- April 3, 2019
- Classification Date
- May 5, 2020
- Termination Date
- July 22, 2022
- Address
- 1099 8th Ave W Suite 106, N/A, Vancouver, N/A, N/A, Canada
Description
Delfi Medical Innovations, VariFit Contour Cuffs, single use adult thigh cuff, single port, single bladder, REF 9-9300-003. Rx Only, Sterile
Delfi Medical Innovations Inc has discovered that tourniquets cuffs in certain lots of the VariFit Sterile Single Use Thigh Cuffs, Single Port, Single Bladder models may have been manufactured with an inadequate perimeter weld on the cuff. Once applied to a patients thigh and inflated to a pressure to occlude blood flow in the thigh distal to the cuff, after a period of time the cuff may leak to the extent that the tourniquet instrument supplying air pressure to the cuff may not be able to adequately control and regulate pressure in the cuff. This may result in some blood flow in the thigh distal to the cuff.
18J17
US: Ohio OUS: Austrailia
59 units