Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android Product Usage: This is a digital product available for download direct to customers from the Apple App Store and the Google Play Store. The total number of downloads is not available. The number of unique users for each affected version from the date of initial distribution through 03/23/2017 is noted in 7a. However, some users will have used all three affected versions as they upgraded from version to version when the versions became available and many others will have tried the app and discontinued use. In the last 30 days, 02/22/2017 03/23/2017, there have been 16,781 unique users in the US with 264 of those seeking bolus advice and 19,268 unique users in the rest of world (excluding Canada) with 415 of those users seeking bolus advice.
Enforcement
- Recall Number
- Z-1899-2017
- Event ID
- 76870
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diabetes Care, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Other
- Report Date
- May 10, 2017
- Initiation Date
- March 14, 2017
- Classification Date
- May 1, 2017
- Termination Date
- January 16, 2018
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025, United States
Description
Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android Product Usage: This is a digital product available for download direct to customers from the Apple App Store and the Google Play Store. The total number of downloads is not available. The number of unique users for each affected version from the date of initial distribution through 03/23/2017 is noted in 7a. However, some users will have used all three affected versions as they upgraded from version to version when the versions became available and many others will have tried the app and discontinued use. In the last 30 days, 02/22/2017 03/23/2017, there have been 16,781 unique users in the US with 264 of those seeking bolus advice and 19,268 unique users in the rest of world (excluding Canada) with 415 of those users seeking bolus advice.
A program error (bug) in the Bolus Advisor feature, which could result in incorrect bolus advice and a potential insulin over-delivery.
Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android
US Nationwide in the states of: AL, AR, AZ, CA, CO, CT, DE, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV
64,324