FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23 566-16130-33
Enforcement
- Recall Number
- Z-1898-2024
- Event ID
- 94443
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- FUJIFILM Healthcare Americas Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 5, 2024
- Initiation Date
- April 11, 2024
- Classification Date
- May 24, 2024
- Address
- 81 Hartwell Ave Ste 300, Lexington, MA, 02421-3160, United States
Description
FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23 566-16130-33
Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury
UDI-DI: 04540217052226 04540217057450 S/N: MP95A9F6A001, MP95A9F74002, MP95A9F74001, MP95AA983002, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA993001, MP95AA996002, MP95AA996001, MP95AA997001, MP95A9F6B001, MP95A9F76001, MP95AA983001, MP95AA988002, MP95AA98A001, MP95AA993002
US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.
18 units