FDA Enforcement Class II Ongoing

FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23 566-16130-33

Recall: Z-1898-2024 · Reported June 5, 2024

Enforcement

Recall Number
Z-1898-2024
Event ID
94443
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
FUJIFILM Healthcare Americas Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 5, 2024
Initiation Date
April 11, 2024
Classification Date
May 24, 2024
Address
81 Hartwell Ave Ste 300, Lexington, MA, 02421-3160, United States

Description

FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23 566-16130-33

Reason

Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury

Code Info

UDI-DI: 04540217052226 04540217057450 S/N: MP95A9F6A001, MP95A9F74002, MP95A9F74001, MP95AA983002, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA993001, MP95AA996002, MP95AA996001, MP95AA997001, MP95A9F6B001, MP95A9F76001, MP95AA983001, MP95AA988002, MP95AA98A001, MP95AA993002

Distribution

US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.

Quantity

18 units