FDA Enforcement Class II Terminated

Adult Divided Cannula end tidal C)2 sampling w/male luer-lock connector. 7 ft, oxygen tube, 7 ft. CO2 tube RX only.For Single Patient Use only. Distributed by Tri-Anium Assembled in Mexico. Usage: Delivery of Anesthesia.

Recall: Z-1898-2012 · Reported July 4, 2012

Enforcement

Recall Number
Z-1898-2012
Event ID
62272
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
SALTER LABS
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
July 4, 2012
Initiation Date
May 30, 2012
Classification Date
June 27, 2012
Termination Date
July 27, 2012
Address
100 Sycamore Rd, Arvin, CA, 93203-2300, United States

Description

Adult Divided Cannula end tidal C)2 sampling w/male luer-lock connector. 7 ft, oxygen tube, 7 ft. CO2 tube RX only.For Single Patient Use only. Distributed by Tri-Anium Assembled in Mexico. Usage: Delivery of Anesthesia.

Reason

A complaint was received, and later verified, that product labeled as SO-1296 with male luer lock connector actually contained product with a female luer lock connector.

Code Info

Salter Part Number SO-1296, Tri-Anium Part numaber: 301-P3600; Lot number: 032612.

Distribution

Nationwide distribution: USA including states of: FL, IN, KS, PA and TX.

Quantity

242 cases