FDA Enforcement Class II Terminated

OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine. REF: MJS-101; MJSR-101

Recall: Z-1896-2021 · Reported June 23, 2021

Enforcement

Recall Number
Z-1896-2021
Event ID
88075
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ossur Americas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 23, 2021
Initiation Date
June 1, 2021
Classification Date
June 17, 2021
Termination Date
June 22, 2023
Address
27051 Towne Centre Dr, Foothill Ranch, CA, 92610-2804, United States

Description

OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine. REF: MJS-101; MJSR-101

Reason

Due to insufficient column strength of the racks for all load conditions, racks associated with height adjustment functionality have been observed deforming, which may lead to increased range of motion in the cervical spine.

Code Info

REF Number: MJS-101; MJSR-101. Lot Numbers: MX180425 through MX200306

Distribution

U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Not provided.

Quantity

10,469 items