FDA Enforcement Class II Ongoing

Signature Laparoscopic Instruments, Dolphin Nose Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5860020.

Recall: Z-1893-2025 · Reported June 11, 2025

Enforcement

Recall Number
Z-1893-2025
Event ID
96755
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Silex Medical, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
June 11, 2025
Initiation Date
January 9, 2024
Classification Date
June 2, 2025
Address
24 Robert Porter Rd Ste B, N/A, Southington, CT, 06489-1158, United States

Description

Signature Laparoscopic Instruments, Dolphin Nose Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5860020.

Reason

Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.

Code Info

Model/REF Number: 5860020. UDI-DI: B33158600200. Lot Number: M28181. Serial Numbers: 0691, 0692.

Distribution

US Nationwide distribution in the states of Florida, North Dakota, Illinois, California, Texas, Virginia, Michigan, Connecticut, Oregon, Iowa, Mississippi, New Jersey, Idaho, West Virginia, Hawaii.

Quantity

2 units