FDA Enforcement Class II Ongoing

Inpeco FlexLab Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf

Recall: Z-1887-2020 · Reported May 13, 2020

Enforcement

Recall Number
Z-1887-2020
Event ID
85429
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Inpeco S.A.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 13, 2020
Initiation Date
April 7, 2020
Classification Date
May 5, 2020
Address
Via San Gottardo 10, N/A, Lugano, N/A, N/A, Switzerland

Description

Inpeco FlexLab Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf

Reason

The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.

Code Info

Serial Numbers: FLX.002, FLX.006, FLX.008, FLX.011, FLX.0121, FLX.0127, FLX.0128, FLX.0129, FLX.0132, FLX.0133, FLX.0139, FLX.014, FLX.0147, FLX.0149, FLX.0150, FLX.0153, FLX.0154, FLX.0160, FLX.0162, FLX.0167, FLX.0168, FLX.0172, FLX.0173, FLX.0174, FLX.0179, FLX.0180, FLX.0182, FLX.0191, FLX.0192, FLX.0193, FLX.0194, FLX.0197, FLX.0198, FLX.0201, FLX.0202, FLX.0203, FLX.0206, FLX.022, FLX.032, FLX.035, FLX.036, FLX.037, FLX.039, FLX.043, FLX.046, FLX.050, FLX.056, FLX.057, FLX.060, FLX.061, FLX.062, FLX.073, FLX.082, FLX.083, FLX.085, FLX.086, FLX.087, FLX.093, FLX.096, FLX.097, FLX.098, FLX.100, SHD.001, SHD.002, SHD.003

Distribution

Worldwide distribution including the following states: Ohio, New York, Virginia, Pennsylvania, Texas, Delaware, California, Florida, Kentucky, Tennessee, Maryland, Rhode Island, Alabama, Massachusetts, Connecticut, Minnesota, Alabama, Montana, Louisiana, New Jersey, Virginia, New Hampshire, Michigan, North Carolina, Washington, Iowa, New Mexico, Nebraska, and Missouri

Quantity

65 units