FDA Enforcement Class II Ongoing

8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. The Stryker Electromagnetic Navigation Unit (8000-010-003) powered by Scopis ENT Software with TGS (8000-021-002), allows the analysis and identification of sinus cells in the complex patient anatomy and planning of the natural drainage pathways through the sinus cavity. During surgery, the planned pathways are overlaid in real-time onto the endoscopic image, providing a unique Scopis augmented reality technology. Guidance of endoscopic instruments may help perform a minimally invasive, accurate and selective surgery.

Recall: Z-1885-2024 · Reported May 29, 2024

Enforcement

Recall Number
Z-1885-2024
Event ID
94456
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Stryker Leibinger GmbH & Co. KG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 29, 2024
Initiation Date
April 4, 2024
Classification Date
May 23, 2024
Address
Botzinger Str. 41, Freiburg Im Breisgau, N/A, Germany

Description

8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. The Stryker Electromagnetic Navigation Unit (8000-010-003) powered by Scopis ENT Software with TGS (8000-021-002), allows the analysis and identification of sinus cells in the complex patient anatomy and planning of the natural drainage pathways through the sinus cavity. During surgery, the planned pathways are overlaid in real-time onto the endoscopic image, providing a unique Scopis augmented reality technology. Guidance of endoscopic instruments may help perform a minimally invasive, accurate and selective surgery.

Reason

Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.

Code Info

Sofware Version: NOVA 3.6.0 RC16, Product Code: 8000-021-002; UDI/DI: 07613327514179; Serial Numbers: 10984, 10110, 10164, 10454, 10762, 10387, 10120, 10252, 10414, 10612, 10408, 10673, 11013, 11032, 10566, 10974, 0310-B, 10326, 10442, 10729, 10681, 10552, 10088, 10476, 10307, 10121, 10371, 0252-B, 10262, 10602, 10613, 10381, 10234, 200774, 10135, 0304-B, 10303, 10257, 10639, 10445, 10684, 10740, 10986, 10413, 10079, 10528, 10565, 10482, 10209, 10574, 10606, 10229, 10214, 10797, 0335-B, 10284, 10465, 10295, 10308, 0282-B, 83B750AD, 0287-B, 10763, 10434, 11025, 10578, 10473, 10596, 0291-B, 10624, 10591, 11092, 10758, 10097, 10294, 0251-B, 10597, 10769, 0309-B, 10765, 0313-B, 0281-B, 10513, 11088, 10057, 10337, 10464, 10621, 10699, 10509, 10421, 10549, 10003, 10499, 10661, 10668, 0292-B, 10416, 0263-B, 10654, 10571, 0329-B, 10934, 10798, 11022, 10610, 10766, 10471, 10437, 10648, 10410, 10403, 10159, 10192, 10372, 10189, 10193, 10642, 10390, 10771, 10202, 10806, 10237, 10296, 0244-B, 0307-B, 10034, 10440, 10704, 10101, 10153, 10098, 10388, 10720, 10943, 10944, 10994, 10695, 10608, 10615, 10054, 10076, 10015, 10089, 10208, 10776, 10780, 11111, 10586, 10759, 10431, 10772, 10376, 10563, 10978, 10989, 10369, 10641, 10796, 10558, 10660, 10071, 10987, 10953, 10982, 10744, 10026, 0325-B, 10959, 11000, 10453, 0332-B, 10055, 10335, 10742, 10649, 10983, 10512, 10958, 10268, 10640, 0280-B, 11075, 11117, 10356, 10429, 10432, 10659, 10677, 10319, 10311, 10332, 10757, 10809, 10523, 10382, 10451, 10495, 10997, 11116, 11018, 10083, 10672, 10775, 10053, 10195, 10330, 10161, 10007, 10328, 10949, 11031, 10781, 10990, 10350, 10750, 10191, 10033, 10056, 10446, 10078, 10516, 10373, 10722, 10136, 10210, 10091, 10255, 10253, 10060, 10427, 10162, 10297, 10731, 10935, 0285-B, 10336, 10760, 10386, 10347, 10286, 10165, 10041, 10456, 10312, 10417, 10500, 10133, 10224, 10031, 10040, 10100, 10784, 10248, 10423, 10215, 10951, 10092, 0260-B, 10185, 10127, 10112, 10790, 10299, 10316, 10435, 10459, 10075, 10479, 10357, 10627, 10732, 10734, 0343-B, 0320-B, 10517, 10455, 10590, 11030, 10383, 10043, 10096, 10789, 10631, 10489, 10791, 10220, 10658, 10348, 10143, 10114, 10625, 10799, 11108, 10103, 10522, 10126, 10066, 10980, 11102, 0318-B, 11005, 10177, 10662, 10166, 10197, 10300, 10406, 10181, 0326-B, 10443, 10331, 10364, 10368, 10592, 10062, 10366, 10349, 10353, 11097, 10155, 10218, 10227, 10230, 10315, 10618, 10021, 10787, 11038, 0271-B, 10365, 10637, 10141, 10525, 10341, 10433, 10696, 0333-B, 10401, 10633, 10211, 10231, 10052, 10794, 10685, 10157, 10213, 10081, 10123, 10090, 10199, 10678, 11020, 10216, 10377, 10424, 10212, 10361, 10059, 10109, 10736, 10344, 10693, 10977, 11004, 10016, 10168, 10171, 10691, 10006, 10458, 10324, 10156, 10190, 10221, 10363, 10475, 0322-B, 0339-B, 10692, 10679, 10082, 10573, 10603, 10267, 10541, 10999, 10169, 10628, 10576, 10345, 10738, 10354, 10238, 0259-B, 10077, 0306-B, 10461, 10172, 10243, 10233, 10327, 0327-B, 11023, 11114, 10450, 10398, 10404, 0303-B, 10400, 10412, 10593, 10072, 10514, 10239, 10069, 10244, 10249, 10761, 10359, 10739, 10352, 10163, 10374, 10698, 10342, 10607, 10635, 10058, 10425, 10419, 10950, 10405, 10422, 10223, 0308-B, 10010, 10594, 10205, 10370, 10474, 10247, 10020, 10174, 10184, 10188, 10128, 11003, 10689, 10701, 10795, 10074, 10266, 10288, 10008, 10203, 10966, 10438, 10955, 10073, 10080, 10333, 10777, 10508, 10325, 10086, 10104, 10469, 10755, 10810, 10671, 10152, 10393, 10657, 10105, 10298, 10524, 10805, 10375, 10402, 10094, 10358, 10498, 10501, 10384, 10391, 10993, 10064, 10175, 10397, 10560, 10570, 10630, 10132, 10579, 10439, 10415, 10306, 10343, 10477, 10323, 10737, 10589, 10304, 10338, 10460, 10598, 10779, 10988, 10346, 11082, 10436, 10362, 10334, 10682, 10653, 10583, 10724, 10725, 10407, 10380, 10409, 10595, 10137, 10656, 10389, 10611, 10638, 11036, 11037, 10577, 10609, 10587, 10626, 10965, 11017, 11112, 10547, 10954, 11029, 11026, 10979, 10985, 11080, 10981, 11099, 10207, 10049, 10580, 10144, 10723.

Distribution

US Nationwide distribution.

Quantity

705 units