Torque Limiting Adapter: Torque Limiting Attachment, Non Sterile, Distributed by Zimmer. PRECIMED, Torque Limiting Adaptor, Non Sterile.. Customer Number/Model: 00-2360-080-00 T10232, T116245 T116245, TLA025-25 TLA025-25, 00-2360-080-05 T10233, MAC21003 T116430, TLA025-30 TLA025-30, MAC21001 T112972, MAC21001 T13621, TLA025-20 TLA025-20.

Recall: Z-1884-2016 · Reported June 8, 2016

Enforcement

Recall Number
Z-1884-2016
Event ID
73948
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Greatbatch Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 8, 2016
Initiation Date
April 26, 2016
Classification Date
June 1, 2016
Termination Date
June 7, 2018
Address
2300 Berkshire Ln N, N/A, Plymouth, MN, 55441-4575, United States

Description

Torque Limiting Adapter: Torque Limiting Attachment, Non Sterile, Distributed by Zimmer. PRECIMED, Torque Limiting Adaptor, Non Sterile.. Customer Number/Model: 00-2360-080-00 T10232, T116245 T116245, TLA025-25 TLA025-25, 00-2360-080-05 T10233, MAC21003 T116430, TLA025-30 TLA025-30, MAC21001 T112972, MAC21001 T13621, TLA025-20 TLA025-20.

Reason

Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.

Code Info

1000207 1001049 1001048 1000698 1000582 1000582 1001334 1001334 1002457 1002456 1002974 1002640 1003175 1003174 1003223 1003223 1003224 1003225 1003548 1003766 1003908 1004435 1004485 1004349 1004903 1005099 1004347 1005530 1005772 1005900 1009013 1008458 1009195 1009195 1009197 1009014 1009198 1009196 1009698 1009199 1009699 1009200 1009700 1011763 1010602 1012455 1012490 1013172 1013173 1013415 1013416 1013921 1013922 1514611 1614605 1714700 1953758 1953809 1953810 2124866 2073128 2116179 2141896 1953758 2150164 2157328 2209891 2302780 2302777 2302781 2302788 2362691 2407040 2407047 2407040 2407040 2898285 2898285 2898289 2971943 2904251 2919512 2938191 2978582 1004138 1004139 1004413 1004578 1004577 1004580 1004579 1005899 1005817 1009701 2324706 2920624 3092836 3092836 1006939 FC-002 FC-002 1012458 6280900001 8027435 8027436 8027434

Distribution

US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.

Quantity

5952