Arrow UltraFlex Intra-Aortic Balloon Catheter Kit, a) REF IAB-06830-U; b) REF IAB-06840-U; c) REF IAB-06850-U

Recall: Z-1879-2024 · Reported June 12, 2024

Enforcement

Recall Number
Z-1879-2024
Event ID
94526
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
ARROW INTERNATIONAL Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 12, 2024
Initiation Date
April 29, 2024
Classification Date
June 3, 2024
Address
3015 Carrington Mill Blvd, N/A, Morrisville, NC, 27560-5437, United States

Description

Arrow UltraFlex Intra-Aortic Balloon Catheter Kit, a) REF IAB-06830-U; b) REF IAB-06840-U; c) REF IAB-06850-U

Reason

Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.

Code Info

a) REF IAB-06830-U, UDI: (01)10801902161939(17)240731(10)18F22H0008, (01)10801902161939(17)241130(10)18F22M0019, (01)10801902161939(17)241231(10)18F23A0027, (01)10801902161939(17)241231(10)18F23A0039, (01)10801902161939(17)250131(10)18F23B0006, (01)10801902161939(17)250131(10)18F23B0020, (01)10801902161939(17)250131(10)18F23B0032, (01)10801902161939(17)250630(10)18F23G0066, (01)10801902161939(17)250731(10)18F23H0059, (01)10801902161939(17)250731(10)18F23H0070, (01)10801902161939(17)250731(10)18F23H0074, (01)10801902161939(17)250831(10)18F23J0046, (01)10801902161939(17)250930(10)18F23K0059, (01)10801902161939(17)251231(10)18F24A0036, (01)10801902161939(17)260331(10)18F24D0004, (01)10801902172898(17)240430(10)18F22E0062, (01)10801902172898(17)240531(10)18F22F0007, (01)10801902172898(17)240531(10)18F22F0017, (01)10801902172898(17)240531(10)18F22F0020, (01)10801902172898(17)240531(10)18F22F0034, (01)10801902172898(17)240531(10)18F22F0045, (01)10801902172898(17)240531(10)18F22F0057, (01)10801902172898(17)240630(10)18F22G0001, (01)10801902172898(17)240630(10)18F22G0004, (01)10801902172898(17)240630(10)18F22G0007, (01)10801902172898(17)240630(10)18F22G0014, (01)10801902172898(17)240731(10)18F22H0002, (01)10801902172898(17)240731(10)18F22H0024, (01)10801902172898(17)240731(10)18F22H0029, (01)10801902172898(17)240731(10)18F22H0034, (01)10801902172898(17)240831(10)18F22J0009, (01)10801902172898(17)250430(10)18F23D0029, (01)10801902172898(17)250430(10)18F23E0037, (01)10801902172898(17)250430(10)18F23E0038, (01)10801902172898(17)250531(10)18F23E0033, (01)10801902172898(17)250531(10)18F23E0057, (01)10801902172898(17)250531(10)18F23F0023, (01)10801902172898(17)250531(10)18F23F0045, (01)10801902172898(17)250630(10)18F23G0025, (01)10801902172898(17)250630(10)18F23G0026, (01)10801902172898(17)250630(10)18F23G0051, (01)10801902172898(17)250731(10)18F23G0067, (01)10801902172898(17)250731(10)18F23H0009, (01)10801902172898(17)250731(10)18F23H0025, (01)10801902172898(17)250731(10)18F23H0028, (01)10801902172898(17)250731(10)18F23H0069, (01)10801902172898(17)250731(10)18F23H0080, (01)10801902172898(17)250831(10)18F23H0011, (01)10801902172898(17)250831(10)18F23J0004, (01)10801902172898(17)250831(10)18F23J0011, (01)10801902172898(17)250831(10)18F23J0015, (01)10801902172898(17)250831(10)18F23J0053, (01)10801902172898(17)250831(10)18F23J0054, (01)10801902172898(17)250831(10)18F23J0055, (01)10801902172898(17)250930(10)18F23J0079, (01)10801902172898(17)250930(10)18F23K0003, (01)10801902172898(17)250930(10)18F23K0005, (01)10801902172898(17)250930(10)18F23K0009, (01)10801902172898(17)250930(10)18F23K0051, (01)10801902172898(17)250930(10)18F23K0061, (01)10801902172898(17)251031(10)18F23L0040, (01)10801902172898(17)251130(10)18F23M0002, (01)10801902172898(17)251130(10)18F23M0044, (01)10801902172898(17)251130(10)18F23M0049, (01)10801902172898(17)251130(10)18F23M0053, (01)10801902172898(17)251130(10)18F23M0054, (01)10801902172898(17)251231(10)18F23M0055, (01)10801902172898(17)251231(10)18F24A0006, (01)10801902172898(17)251231(10)18F24A0012, (01)10801902172898(17)251231(10)18F24A0013, (01)10801902172898(17)251231(10)18F24A0021, (01)10801902172898(17)260131(10)18F24B0014, (01)10801902172898(17)260228(10)18F24C0004, (01)10801902172898(17)260228(10)18F24C0005, (01)10801902172898(17)260331(10)18F24D0033, (01)10801902172898(17)260331(10)18F24D0036, (01)10801902172898(17)260331(10)18F24D0043, (01)10801902172898(17)260331(10)18F24D0048, (01)10801902172898(17)260331(10)18F24D0086; b) REF IAB-06840-U, UDI: (01)10801902161946(17)240430(10)18F22E0023, (01)10801902161946(17)240531(10)18F22F0009, (01)10801902161946(17)240531(10)18F22F0019, (01)10801902161946(17)240731(10)18F22H0014, (01)10801902161946(17)240731(10)18F22H0028, (01)10801902161946(17)240731(10)18F22H0036, (01)10801902161946(17)241130(10)18F22M0015, (01)10801902161946(17)241231(10)18F23A0019, (01)10801902161946(17)241231(10)18F23A0037, (01)10801902161946(17)241231(10)18F23A0054, (01)10801902161946(17)241231(10)18F23A0055, (01)10801902161946(17)241231(10)18F23A0061, (01)10801902161946(17)250131(10)18F23A0066, (01)10801902161946(17)250131(10)18F23A0067, (01)10801902161946(17)250228(10)18F23A0025, (01)10801902161946(17)250228(10)18F23B0018, (01)10801902161946(17)250331(10)18F23D0018, (01)10801902161946(17)250430(10)18F23E0013, (01)10801902161946(17)250430(10)18F23E0015, (01)10801902161946(17)250630(10)18F23G0002, (01)10801902161946(17)250630(10)18F23G0003, (01)10801902161946(17)250630(10)18F23G0011, (01)10801902161946(17)250731(10)18F23G0063, (01)10801902161946(17)250731(10)18F23H0034, (01)10801902161946(17)250731(10)18F23H0035, (01)10801902161946(17)250731(10)18F23H0038, (01)10801902161946(17)250731(10)18F23H0053, (01)10801902161946(17)250731(10)18F23H0058, (01)10801902161946(17)250731(10)18F23H0071, (01)10801902161946(17)250831(10)18F23J0006, (01)10801902161946(17)250831(10)18F23J0008, (01)10801902161946(17)250831(10)18F23J0014, (01)10801902161946(17)250831(10)18F23J0018, (01)10801902161946(17)250831(10)18F23J0035, (01)10801902161946(17)250831(10)18F23J0042, (01)10801902161946(17)250930(10)18F23K0028, (01)10801902161946(17)260131(10)18F24B0025, (01)10801902161946(17)260131(10)18F24B0026, (01)10801902161946(17)260228(10)18F24C0151, (01)10801902161946(17)260331(10)18F24C0171, (01)10801902161946(17)260331(10)18F24C0177, (01)10801902161946(17)260331(10)18F24C0178, (01)10801902161946(17)260331(10)18F24C0179, (01)10801902161946(17)260331(10)18F24C0180, (01)10801902161946(17)260331(10)18F24C0195, (01)10801902161946(17)260331(10)18F24D0014, (01)10801902161946(17)260331(10)18F24D0017, (01)10801902161946(17)260331(10)18F24D0030, (01)10801902172904(17)240430(10)18F22E0001, (01)10801902172904(17)240430(10)18F22E0009, (01)10801902172904(17)240430(10)18F22E0014, (01)10801902172904(17)240430(10)18F22E0020, (01)10801902172904(17)240430(10)18F22E0036, (01)10801902172904(17)240430(10)18F22E0037, (01)10801902172904(17)240430(10)18F22E0042, (01)10801902172904(17)240430(10)18F22E0046, (01)10801902172904(17)240430(10)18F22E0048, (01)10801902172904(17)240430(10)18F22E0049, (01)10801902172904(17)240430(10)18F22E0061, (01)10801902172904(17)240430(10)18F22E0065, (01)10801902172904(17)240430(10)18F22E0066, (01)10801902172904(17)240531(10)18F22F0002, (01)10801902172904(17)240531(10)18F22F0010, (01)10801902172904(17)240531(10)18F22F0016, (01)10801902172904(17)240531(10)18F22F0021, (01)10801902172904(17)240531(10)18F22F0022, (01)10801902172904(17)240531(10)18F22F0037, (01)10801902172904(17)240531(10)18F22F0043, (01)10801902172904(17)240531(10)18F22F0047, (01)10801902172904(17)240531(10)18F22F0050, (01)10801902172904(17)240531(10)18F22F0051, (01)10801902172904(17)240531(10)18F22F0054, (01)10801902172904(17)240531(10)18F22F0065, (01)10801902172904(17)240531(10)18F22F0066, (01)10801902172904(17)240531(10)18F22F0067, (01)10801902172904(17)240531(10)18F22H0026, (01)10801902172904(17)240630(10)18F22G0067, (01)10801902172904(17)240731(10)18F22H0004, (01)10801902172904(17)240731(10)18F22H0010, (01)10801902172904(17)240731(10)18F22H0030, (01)10801902172904(17)240731(10)18F22H0038, (01)10801902172904(17)240731(10)18F22H0039, (01)10801902172904(17)240930(10)18F22K0004, (01)10801902172904(17)250228(10)18F23C0017, (01)10801902172904(17)250228(10)18F23C0027, (01)10801902172904(17)250228(10)18F23C0044, (01)10801902172904(17)250331(10)18F23D0015, (01)10801902172904(17)250331(10)18R23G0002, (01)10801902172904(17)250430(10)18F23D0027, (01)10801902172904(17)250430(10)18F23D0028, (01)10801902172904(17)250430(10)18F23E0014, (01)10801902172904(17)250430(10)18F23E0028, (01)10801902172904(17)250430(10)18F23E0035, (01)10801902172904(17)250430(10)18F23E0036, (01)10801902172904(17)250430(10)18F23E0050, (01)10801902172904(17)250430(10)18F23E0051, (01)10801902172904(17)250430(10)18F23E0052, (01)10801902172904(17)250531(10)18F23E0055, (01)10801902172904(17)250531(10)18F23F0003, (01)10801902172904(17)250531(10)18F23F0006, (01)10801902172904(17)250531(10)18F23F0007, (01)10801902172904(17)250531(10)18F23F0014, (01)10801902172904(17)250531(10)18F23F0015, (01)10801902172904(17)250531(10)18F23F0024, (01)10801902172904(17)250531(10)18F23F0025, (01)10801902172904(17)250531(10)18F23F0048, (01)10801902172904(17)250531(10)18F23F0052, (01)10801902172904(17)250531(10)18F23F0054, (01)10801902172904(17)250531(10)18F23F0059, (01)10801902172904(17)250630(10)18F23G0005, (01)10801902172904(17)250630(10)18F23G0007, (01)10801902172904(17)250630(10)18F23G0014, (01)10801902172904(17)250630(10)18F23G0028, (01)10801902172904(17)250630(10)18F23G0050, (01)10801902172904(17)250630(10)18F23G0053, (01)10801902172904(17)250731(10)18F23G0064, (01)10801902172904(17)250731(10)18F23G0065, (01)10801902172904(17)250731(10)18F23H0057, (01)10801902172904(17)250731(10)18F23H0075, (01)10801902172904(17)250831(10)18F23H0033, (01)10801902172904(17)250831(10)18F23J0019, (01)10801902172904(17)250831(10)18F23J0041, (01)10801902172904(17)250831(10)18F23J0050, (01)10801902172904(17)250930(10)18F23K0017, (01)10801902172904(17)250930(10)18F23K0018, (01)10801902172904(17)250930(10)18F23K0019, (01)10801902172904(17)250930(10)18F23K0027, (01)10801902172904(17)250930(10)18F23K0040, (01)10801902172904(17)250930(10)18F23K0045, (01)10801902172904(17)250930(10)18F23K0057, (01)10801902172904(17)251031(10)18F23K0060, (01)10801902172904(17)251031(10)18F23L0006, (01)10801902172904(17)251031(10)18F23L0029, (01)10801902172904(17)251031(10)18F23L0030, (01)10801902172904(17)251130(10)18F23M0019, (01)10801902172904(17)251130(10)18F23M0020, (01)10801902172904(17)251130(10)18F23M0037, (01)10801902172904(17)251130(10)18F23M0039, (01)10801902172904(17)251130(10)18F23M0040, (01)10801902172904(17)251130(10)18F23M0041, (01)10801902172904(17)251231(10)18F23M0064, (01)10801902172904(17)251231(10)18F23M0065, (01)10801902172904(17)251231(10)18F24A0002, (01)10801902172904(17)251231(10)18F24A0003, (01)10801902172904(17)251231(10)18F24A0004, (01)10801902172904(17)251231(10)18F24A0005, (01)10801902172904(17)251231(10)18F24A0014, (01)10801902172904(17)251231(10)18F24A0016, (01)10801902172904(17)251231(10)18F24A0022, (01)10801902172904(17)251231(10)18F24A0023, (01)10801902172904(17)251231(10)18F24A0024, (01)10801902172904(17)251231(10)18F24A0038, (01)10801902172904(17)260131(10)18F24B0006, (01)10801902172904(17)260131(10)18F24B0007, (01)10801902172904(17)260131(10)18F24B0023, (01)10801902172904(17)260131(10)18F24B0024, (01)10801902172904(17)260131(10)18F24B0027, (01)10801902172904(17)260131(10)18F24B0028, (01)10801902172904(17)260131(10)18F24B0045, (01)10801902172904(17)260131(10)18F24B0046, (01)10801902172904(17)260131(10)18F24B0047, (01)10801902172904(17)260131(10)18F24B0068, (01)10801902172904(17)260131(10)18F24B0069, (01)10801902172904(17)260228(10)18F24B0067, (01)10801902172904(17)260228(10)18F24B0070, (01)10801902172904(17)260228(10)18F24B0071, (01)10801902172904(17)260228(10)18F24C0006, (01)10801902172904(17)260228(10)18F24C0007, (01)10801902172904(17)260228(10)18F24C0028, (01)10801902172904(17)260228(10)18F24C0029, (01)10801902172904(17)260228(10)18F24C0030, (01)10801902172904(17)260228(10)18F24C0038, (01)10801902172904(17)260228(10)18F24C0092, (01)10801902172904(17)260228(10)18F24C0120, (01)10801902172904(17)260228(10)18F24C0121, (01)10801902172904(17)260228(10)18F24C0141; c) REF IAB-06850-U, UDI: (01)10801902144192(17)240831(10)18F22J0020(13)220916, (01)10801902144192(17)240831(10)18F22J0038(13)220928, (01)10801902144192(17)240831(10)18F22J0043(13)220930, (01)10801902144192(17)240930(10)18F22K0001(13)221003, (01)10801902144192(17)240930(10)18F22K0003(13)221004, (01)10801902144192(17)240930(10)18F22K0006(13)221006, (01)10801902144192(17)240930(10)18F22K0011(13)221007, (01)10801902144192(17)240930(10)18F22K0012(13)221007, (01)10801902144192(17)240930(10)18F22K0013(13)221007, (01)10801902144192(17)241231(10)18F23A0075(13)230131, (01)10801902144192(17)250131(10)18F23B0005(13)230203, (01)10801902144192(17)250131(10)18F23B0010(13)230206, (01)10801902144192(17)250131(10)18F23B0029(13)230220, (01)10801902144192(17)250131(10)18F23B0030(13)230220, (01)10801902144192(17)250228(10)18F23C0007(13)230307, (01)10801902144192(17)250228(10)18F23C0008(13)230307, (01)10801902144192(17)250228(10)18F23C0014(13)230310, (01)10801902144192(17)250228(10)18F23C0015(13)230310, (01)10801902144192(17)250531(10)18F23F0016(13)230609, (01)10801902144192(17)250531(10)18F23F0017(13)230609, (01)10801902144192(17)250831(10)18F23J0043(13)230921, (01)10801902144192(17)250831(10)18F23J0044(13)230923, (01)10801902144192(17)250930(10)18F23J0078(13)231031, (01)10801902144192(17)250930(10)18F23K0015(13)231031, (01)10801902144192(17)250930(10)18F23K0024(13)231014, (01)10801902144192(17)251031(10)18F23L0045(13)231130, (01)10801902144192(17)251130(10)18F23L0052(13)231205, (01)10801902144192(17)251130(10)18F23L0058(13)231205, (01)10801902144192(17)260131(10)18F24B0011(13)240209, (01)10801902144192(17)260131(10)18F24B0012(13)240209, (01)10801902172911(17)250831(10)18F23H0061(13)230911, (01)10801902172911(17)260131(10)18F24B0010(13)240207

Distribution

Worldwide distribution. US nationwide including Puerto Rico, AU, BA, BE, BG, BN, CA, CL, CN, CO, CY, CZ, DE, EC, ES, FR, GB, GP, GR, HU, HK, IE, IL, IN, IQ, IT, JO, JP, KW, KZ, LB, LU, MA, MN, MK, MX, MY, NL, OM, PA, PF, PH, PL, PT, QA, RE, RO, RS, SA, SE, SG, SI, SK, TH, TR, TW, UZ, ZA, and UAE.

Quantity

44,807 units