FDA Enforcement Class II Terminated

Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343, 54344, and 100105-01.

Recall: Z-1878-2016 · Reported June 8, 2016

Enforcement

Recall Number
Z-1878-2016
Event ID
74004
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Natus Medical Incorporated
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
June 8, 2016
Initiation Date
April 22, 2016
Classification Date
June 1, 2016
Termination Date
December 16, 2016
Address
5900 1st Ave S, N/A, Seattle, WA, 98108-3248, United States

Description

Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343, 54344, and 100105-01.

Reason

The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down during use, contaminating the interior chamber of the Sterile Drier and its contents.

Code Info

Filter Serial Numbers: Filter SN: 2522893, 2522897, 2522898, 2522899, 2522900, 2522903, 2522914, 2522915, 2522916, 2522917, 2522918, 2522919, 2522920, 2522921, 2522922, 2522923, 2522924, 2522925, 2522926, 2522930, and 2522931.

Distribution

distributed in the US including AZ, CA, FL, IL, KS, MA, MD, MN, NC, NH, NY, OH, OK, PA, VA, WA, WA DC,and WI and in Canada and Singapore.

Quantity

21 filters