Sonopet Ultrasonic Aspirator Console (Console 110V with Pedal & Pole) Product Usage: The Sonopet Console is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft and hard tissue is desirable, including neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, plastic and reconstructive surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and thoracoscopic surgery.
Enforcement
- Recall Number
- Z-1878-2013
- Event ID
- 65486
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Instruments Div. of Stryker Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 14, 2013
- Initiation Date
- June 24, 2013
- Classification Date
- August 5, 2013
- Termination Date
- January 30, 2014
- Address
- 4100 E Milham Ave, N/A, Portage, MI, 49002-9704, United States
Description
Sonopet Ultrasonic Aspirator Console (Console 110V with Pedal & Pole) Product Usage: The Sonopet Console is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft and hard tissue is desirable, including neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, plastic and reconstructive surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and thoracoscopic surgery.
A supplied component exhibiting corrosion was built into the power supply boards of the Sonopet Consoles listed above. The issue is limited to 2 lots of the supplied component. The presence of the corrosion on the relay of the power supply board can potentially lead to a conduction defect which could cause the irrigation pump to stop working. The purpose of the irrigation pump is to supply sali
5450-850-000 (Console 110V with Pedal & Pole) S/N 1131300134, 1131900074, 1200500134, 1203200244, 1210100114, 1131300144, 1131900084, 1200500144, 1203200254, 1210100124, 1131300154, 1131900094, 1200500154, 1203200274, 1210100134, 1131300164, 1131900104, 1200500164, 1203700954, 1210100144, 1131300174, 1131900114, 1200500174, 1203700964, 1210100154, 1131300184, 1131900124, 1200500184, 1203700974, 1210700014, 1131300194, 1132800114, 1200500194, 1203900024, 1210700024, 1131300204, 1132800124, 1200500204, 1204000044, 1210700034, 1131300214, 1132800134, 1201300494, 1204000054, 1210700044, 1131300224, 1132800144, 1201300544, 1207400154, 1210700054, 1131300234, 1132800154, 1201300754, 1207400164, 1213800114, 1131300244, 1132800164, 1201300764, 1207400174, 1213800124, 1131900014, 1132800174, 1201300774, 1207400184, 1213800134, 1131900024, 1132800184, 1201300784, 1208300014, 1213800144, 1131900034, 1132800194, 1201300794, 1208300024, 1213800154, 1131900044, 1132800204, 1203200214, 1208300034, 1131900054, 1200500114, 1203200224, 1208300044, 1131900064, 1200500124, 1203200234, 1208300054.
Worldwide - USA Nationwide including the states of CA, CO, FL, HI, IA, IL, KS, KY, LA, MD, MI, MN, MS, NC, NH, NJ, NV, NY, OH, PA, TX, VA, ITALY, and the countries of AUSTRALIA, CANADA, UNITED KINGDOM, NETHERLANDS, INDIA, JAPAN, NEW ZEALAND, SWITZERLAND, HONG KONG, and TAIWAN.
215 units total