FDA Enforcement Class II Terminated

Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in normal and neoplastic tissues.

Recall: Z-1877-2016 · Reported June 8, 2016

Enforcement

Recall Number
Z-1877-2016
Event ID
73802
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Leica Microsystems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 8, 2016
Initiation Date
April 5, 2016
Classification Date
June 1, 2016
Termination Date
November 30, 2017
Address
1700 Leider Ln, N/A, Buffalo Grove, IL, 60089-6622, United States

Description

Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in normal and neoplastic tissues.

Reason

Testing of retained units indicated that Lot numbers of CDX2 indicated are not stable up to the expiry date on the product labeling.

Code Info

1ml size Novocastr Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) with lot numbers 6032734, 6033766, 6037153 and 6039088.

Distribution

Distributed in the following states AR, CA, CT, FL, ID, IL, IN, MT, OH, PA, TN and TX, and in Canada.

Quantity

90 Units