Laerdal Compact Suction Unit ¿ 4 under the following labels: 1) 300 ml (3.3 lbs) canister version - CATALOG #(s): 88006101 - LCSU 4, 300 ml Complete unit (US), 88006103 - LCSU 4, 300 ml Complete unit (UK), 88006140 - LCSU 4, 300 ml Complete unit (EUR), 881151 - Replacement unit, 88006201 - LCSU 4, 300 ml, RTCA Complete unit (US), 88006203 - LCSU 4, 300 ml, RTCA Complete unit (UK), 88006240 - LCSU 4, 300 ml, RTCA Complete unit (EUR) and 881152 - Replacement Unit, RTCA; 2) 88005101 - LCSU 4, 800 ml Complete unit (US), 88005103 - LCSU 4, 800 ml Complete unit (UK), ¿ 88005140 - LCSU 4, 800 ml Complete unit (EUR), 88005201 - LCSU 4, 800 ml, RTCA Complete unit (US), 88005203 - LCSU 4, 800 ml, RTCA Complete unit (UK), 88005240 - LCSU 4, 800 ml, RTCA Complete unit (EUR) & 2) 800 ml (4.3 lbs) canister version - CATALOG #(s): 88005101, 88005201 (RTCA)
Enforcement
- Recall Number
- Z-1873-2017
- Event ID
- 76744
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Laerdal Medical Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 26, 2017
- Initiation Date
- January 8, 2017
- Classification Date
- April 20, 2017
- Termination Date
- May 25, 2018
- Address
- PO Box 1840, 167 Myers Corners Rd, Wappingers Falls, NY, 12590-3869, United States
Description
Laerdal Compact Suction Unit ¿ 4 under the following labels: 1) 300 ml (3.3 lbs) canister version - CATALOG #(s): 88006101 - LCSU 4, 300 ml Complete unit (US), 88006103 - LCSU 4, 300 ml Complete unit (UK), 88006140 - LCSU 4, 300 ml Complete unit (EUR), 881151 - Replacement unit, 88006201 - LCSU 4, 300 ml, RTCA Complete unit (US), 88006203 - LCSU 4, 300 ml, RTCA Complete unit (UK), 88006240 - LCSU 4, 300 ml, RTCA Complete unit (EUR) and 881152 - Replacement Unit, RTCA; 2) 88005101 - LCSU 4, 800 ml Complete unit (US), 88005103 - LCSU 4, 800 ml Complete unit (UK), ¿ 88005140 - LCSU 4, 800 ml Complete unit (EUR), 88005201 - LCSU 4, 800 ml, RTCA Complete unit (US), 88005203 - LCSU 4, 800 ml, RTCA Complete unit (UK), 88005240 - LCSU 4, 800 ml, RTCA Complete unit (EUR) & 2) 800 ml (4.3 lbs) canister version - CATALOG #(s): 88005101, 88005201 (RTCA)
The On/Off functionality in the LCSU 4 is controlled by electronic components, one of which might be damaged.
Affected units are labelled with Manufacturing Dates in the range 20150501 to 20161028.
Nationwide
4,377 units