FDA Enforcement Class II Terminated

Ergon III Series Skyboom Monitor Bracket. Part # 3FCM1 Flatscreen Bracket, single monitor mount, 27inch max. screen size

Recall: Z-1871-2016 · Reported June 8, 2016

Enforcement

Recall Number
Z-1871-2016
Event ID
73704
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Skytron, Div. The KMW Group, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
June 8, 2016
Initiation Date
March 28, 2016
Classification Date
May 31, 2016
Termination Date
July 25, 2017
Address
5085 Corporate Exchange Blvd SE, N/A, Grand Rapids, MI, 49512-5515, United States

Description

Ergon III Series Skyboom Monitor Bracket. Part # 3FCM1 Flatscreen Bracket, single monitor mount, 27inch max. screen size

Reason

Potential for the weld to fail at the pivot point of the monitor bracket. This weld failure could cause the display interface and bracket to fall from its mounted location, resulting in injury.

Code Info

Model Number(s): 3FCM1; Manufactured Date: September 2015; Serial Numbers 1507T719-00-0-0074, 1507T719-00-0-0091, 1507T719-00-0-0095, 1507T719-00-0-0096, 1507T719-00-0-0102, 1507T719-00-0-0107, 1508T719-00-0-0129, 1507T719-00-0-0087, 1507T719-00-0-0069, 1507T719-00-0-0103, 1507T719-00-0-0104, 1507T719-00-0-0066, 1507T719-00-0-0072, 1507T719-00-0-0073, 1507T719-00-0-0075, 1507T719-00-0-0088, 1507T719-00-0-0101, 1507T719-00-0-0108, 1507T719-00-0-0112, 1507T719-00-0-0113, 1507T719-00-0-0090, 1507T719-00-0-0097, 1507T719-00-0-0099, and 1507T719-00-0-0100

Distribution

Worldwide distribution. US states: VA, VT, UT, GA, NJ, and IL; Japan.

Quantity

24