FDA Enforcement Class II Terminated

Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A FacetFuse Drill Guides were designed to provide a concentric pathway for the drills and taps to access the bone while protecting the soft tissue from the cutting geometry of the drills/taps. In addition the drill guide is used as a reference to measure depth that the drills and taps are protruding into bone, which is measured by depth markings at the proximal end of the drills/taps relative to the proximal end of the drill guide. These devices were used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients.

Recall: Z-1871-2013 · Reported August 14, 2013

Enforcement

Recall Number
Z-1871-2013
Event ID
64985
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
SpineFrontier, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 14, 2013
Initiation Date
November 9, 2012
Classification Date
August 2, 2013
Termination Date
June 5, 2015
Address
500 Cummings Ctr Ste 3500, N/A, Beverly, MA, 01915-6516, United States

Description

Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A FacetFuse Drill Guides were designed to provide a concentric pathway for the drills and taps to access the bone while protecting the soft tissue from the cutting geometry of the drills/taps. In addition the drill guide is used as a reference to measure depth that the drills and taps are protruding into bone, which is measured by depth markings at the proximal end of the drills/taps relative to the proximal end of the drill guide. These devices were used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients.

Reason

Prior to use in surgery, the Drill Guides were presenting noticeable degrees of denting and/or bending. As a result, drills and taps could not be directed down guides during surgery due to mechanical interference.

Code Info

2520301, 2520307

Distribution

US Distribution including Puerto Rico and the states of CO, KS, MA and TX.

Quantity

13 drill guides