FDA Enforcement Class II Terminated

DriSate¿ Product Code DR-145 The acid concentrate powders and liquids are indicated for use in acute and chronic hemodialysis and to be used with the appropriate hemodialysis machine/monitor and bicarbonate concentrate.

Recall: Z-1869-2018 · Reported May 23, 2018

Enforcement

Recall Number
Z-1869-2018
Event ID
79878
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Rockwell Medical, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
May 23, 2018
Initiation Date
April 7, 2018
Classification Date
May 14, 2018
Termination Date
October 1, 2018
Address
30142 S. Wixom Rd., Wixom, MI, 48393-3440, United States

Description

DriSate¿ Product Code DR-145 The acid concentrate powders and liquids are indicated for use in acute and chronic hemodialysis and to be used with the appropriate hemodialysis machine/monitor and bicarbonate concentrate.

Reason

A subassembly component (dextrose bag) within this case product contained small amounts of sodium bicarbonate from a prior batch.

Code Info

Lot 28562-0220171

Distribution

US Distribution to the state of NY

Quantity

139