FDA Enforcement Class II Terminated

Thermo Scientific General Purpose Dynamic Bath Precision 10L GP Bath Model Number: TSGP10

Recall: Z-1868-2019 · Reported July 3, 2019

Enforcement

Recall Number
Z-1868-2019
Event ID
82903
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Thermo Fisher Scientific (Asheville) LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
July 3, 2019
Initiation Date
April 2, 2019
Classification Date
June 21, 2019
Termination Date
July 23, 2020
Address
401 Mill Creek Rd, N/A, Marietta, OH, 45750-4304, United States

Description

Thermo Scientific General Purpose Dynamic Bath Precision 10L GP Bath Model Number: TSGP10

Reason

Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.

Code Info

Serial Numbers: 300246811 300246812 300246814 300246815 300247403 300247404 300247406 300247407 300247409 300247789 300247791 300247792

Distribution

CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia

Quantity

12 units