Z-800W Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.

Recall: Z-1867-2025 · Reported June 18, 2025

Enforcement

Recall Number
Z-1867-2025
Event ID
96796
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Zyno Medical LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
June 18, 2025
Initiation Date
May 7, 2025
Classification Date
June 6, 2025
Address
177 Pine St, Natick, MA, 01760-1331, United States

Description

Z-800W Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.

Reason

Unreleased software versions were installed on distributed devices without verification or validation.

Code Info

Model Number: Z800W; UDI-DI; 00814371020020; Serial No. 700063; 700163; 700203; 700402; 700488; 700093; 700169; 700210; 700428; 700495; 700104; 700172; 700214; 700431; 700503; 700111; 700188; 700225; 700440; 700510; 700120; 700189; 700230; 700449; 700513; 700135; 700190; 700285; 700460; 700541; 700145; 700192; 700375; 700471; 700481; 700150; 700195; 700391; 700472; 700397; 700159; 700199; 700393; 700478; 700201; 700162.

Distribution

US Nationwide

Quantity

613 units