FDA Enforcement
Class I
Ongoing
Z-800W Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.
Recall: Z-1867-2025
·
Reported June 18, 2025
Enforcement
- Recall Number
- Z-1867-2025
- Event ID
- 96796
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Zyno Medical LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- June 18, 2025
- Initiation Date
- May 7, 2025
- Classification Date
- June 6, 2025
- Address
- 177 Pine St, Natick, MA, 01760-1331, United States
Description
Z-800W Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.
Reason
Unreleased software versions were installed on distributed devices without verification or validation.
Code Info
Model Number: Z800W; UDI-DI; 00814371020020; Serial No. 700063; 700163; 700203; 700402; 700488; 700093; 700169; 700210; 700428; 700495; 700104; 700172; 700214; 700431; 700503; 700111; 700188; 700225; 700440; 700510; 700120; 700189; 700230; 700449; 700513; 700135; 700190; 700285; 700460; 700541; 700145; 700192; 700375; 700471; 700481; 700150; 700195; 700391; 700472; 700397; 700159; 700199; 700393; 700478; 700201; 700162.
Distribution
US Nationwide
Quantity
613 units