FDA Enforcement
Class II
Terminated
Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 20L GP Bath Model Number: FSGPD20
Recall: Z-1867-2019
·
Reported July 3, 2019
Enforcement
- Recall Number
- Z-1867-2019
- Event ID
- 82903
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Thermo Fisher Scientific (Asheville) LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- July 3, 2019
- Initiation Date
- April 2, 2019
- Classification Date
- June 21, 2019
- Termination Date
- July 23, 2020
- Address
- 401 Mill Creek Rd, N/A, Marietta, OH, 45750-4304, United States
Description
Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 20L GP Bath Model Number: FSGPD20
Reason
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Code Info
Serial Numbers: 300247790
Distribution
CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia
Quantity
1 unit