BCS(R) Behring Coagulation System, Device catalog No. OVIO03 Product Usage: Multipurpose system for In Vitro coagulation studies
Enforcement
- Recall Number
- Z-1867-2012
- Event ID
- 61768
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 4, 2012
- Initiation Date
- December 1, 2010
- Classification Date
- June 25, 2012
- Termination Date
- March 5, 2014
- Address
- 500 GBC Drive, Mailstop 514, PO BOX 6101, Newark, DE, 19714-6101, United States
Description
BCS(R) Behring Coagulation System, Device catalog No. OVIO03 Product Usage: Multipurpose system for In Vitro coagulation studies
Firm has confirmed that if the Cleaner SCS vial becomes empty after processing a test for Antithrombin assay using the INNOVANCE(R) Antithrombin kit, and the test immediately following is an APTT based clotting assay several results may be shortened.
Device catalog No. OVIO03 (not lot specific)
Worldwide Distribution - US (nationwide) including the states of: AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IO, KS, KY, LA, MD, MA, MI, MS, MN, MT, NE, NH, NJ, NY, NC, ND, OH, OK, OR, PA, TN, TX, UT, VA, WA, WV, and WI and the countries of: Australia, France, Canary Islands, Spain, Belgium, Germany, Switzerland, Netherlands, Italy, Guadeloupe, Ceuta, Austria, Slovenia, Croatia, Saudi Arabia, Turkey, Greece, and San Marino.