FDA Enforcement Class II Ongoing

Electrostatic Filter Angled, Item Code 350S19006 - Product Usage: indicated for single use on anesthetized patients and respiratory care patients who require a breathing circuit with ISO standard 15mm or 22 mm connectors.

Recall: Z-1866-2021 · Reported June 23, 2021

Enforcement

Recall Number
Z-1866-2021
Event ID
87925
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Covidien, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 23, 2021
Initiation Date
April 29, 2021
Classification Date
June 16, 2021
Address
60 Middletown Ave, North Haven, CT, 06473-3908, United States

Description

Electrostatic Filter Angled, Item Code 350S19006 - Product Usage: indicated for single use on anesthetized patients and respiratory care patients who require a breathing circuit with ISO standard 15mm or 22 mm connectors.

Reason

The company was notified by their supplier of sterilization services for airway products in Italy of potential deviations from validated parameters for ethylene oxide sterilization.

Code Info

GTIN Primary Package: 40884522012981 GTIN Secondary Package: 30884522012984 Lot Numbers: 20G1304FAX 20G1305FAX 20L0026FAX 20L0027FAX 20L0410FAX 20L0411FAX 20L0412FAX

Distribution

Worldwide distribution - US Nationwide distribution in the states of DE, KY, NY, TX, VA, VT, and WA. Global Distribution.

Quantity

32625