FDA Enforcement
Class II
Terminated
BD Viper LT software, version 3.00H
Recall: Z-1866-2017
·
Reported April 26, 2017
Enforcement
- Recall Number
- Z-1866-2017
- Event ID
- 76895
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Bd Diagnostic
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 26, 2017
- Initiation Date
- December 14, 2015
- Classification Date
- April 19, 2017
- Termination Date
- April 19, 2017
- Address
- 54 Loveton Circle Mc 912, N/A, Sparks, MD, 21152, United States
Description
BD Viper LT software, version 3.00H
Reason
BD has confirmed reports that lower than expected RFU values were generated on customer BD Viper LT Systems after the installation of software version 3.00H. As RFU values decrease, results approach the assay cutoff threshold which has the potential to impact test results.
Code Info
Catalog Number 443157 443153 443155 443156 442947 443902 Version Number 3.00H 3.00H
Distribution
2 US customers, 0 Ex-US regions, and 0 distributors 0.
Quantity
2