FDA Enforcement Class II Terminated

BD Viper LT software, version 3.00H

Recall: Z-1866-2017 · Reported April 26, 2017

Enforcement

Recall Number
Z-1866-2017
Event ID
76895
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Bd Diagnostic
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 26, 2017
Initiation Date
December 14, 2015
Classification Date
April 19, 2017
Termination Date
April 19, 2017
Address
54 Loveton Circle Mc 912, N/A, Sparks, MD, 21152, United States

Description

BD Viper LT software, version 3.00H

Reason

BD has confirmed reports that lower than expected RFU values were generated on customer BD Viper LT Systems after the installation of software version 3.00H. As RFU values decrease, results approach the assay cutoff threshold which has the potential to impact test results.

Code Info

Catalog Number 443157 443153 443155 443156 442947 443902 Version Number 3.00H 3.00H

Distribution

2 US customers, 0 Ex-US regions, and 0 distributors 0.

Quantity