FDA Enforcement
Class II
Terminated
Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 5L GP Bath Model Number: FSGPD05
Recall: Z-1865-2019
·
Reported July 3, 2019
Enforcement
- Recall Number
- Z-1865-2019
- Event ID
- 82903
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Thermo Fisher Scientific (Asheville) LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- July 3, 2019
- Initiation Date
- April 2, 2019
- Classification Date
- June 21, 2019
- Termination Date
- July 23, 2020
- Address
- 401 Mill Creek Rd, N/A, Marietta, OH, 45750-4304, United States
Description
Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 5L GP Bath Model Number: FSGPD05
Reason
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Code Info
Serial Numbers: 30024679S 300246796 300246797 300246798 300246799 300247393 300247394 30024739S 300247396 3002477S8
Distribution
CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia
Quantity
10 units