FDA Enforcement Class II Terminated

Stryker Rose Bed Model (FL14E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone

Recall: Z-1861-2012 · Reported July 4, 2012

Enforcement

Recall Number
Z-1861-2012
Event ID
61712
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Medical Division of Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 4, 2012
Initiation Date
May 17, 2012
Classification Date
June 22, 2012
Termination Date
May 12, 2014
Address
3800 East Centre Ave., Portage, MI, 49002, United States

Description

Stryker Rose Bed Model (FL14E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone

Reason

An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac

Code Info

Model FL14E: Serial Number QB22371, QB22370, QB22369, QB22368, QB22367, QB21200, QB18330, QB17733, B22524, B22523, B22522, B22521, B22506, B22505, B22504, B22503, B22502, B22501, B22500, B22446, B22445, B22444, B22443, B22442, B22441, B22440, B22439, B22438, B22437, B22436, B22435, B22427, B22426, B22425, B22424, B22423, B22422, B22421, B22420, B22419, B22373, B22372, B22308, B22307, B22306, B22305, B22304, B22303, B22302, B22301, B22300, B22299, B22298, B22297, B22296, B22295, B22294, B22293, B22292, B22291, B22290, B22289, B22247, B22246, B22245, B22244, B22243, B22242, B22241, B22240, B22239, B22238, B22237, B22236, B22235, B22234, B22233, B22232, B22231, B22230, B22229, B22228, B22227, B22226, B22225, B22224, B22223, B22222, B22221, B22220, B22219, B22218, B22217, B22216, B22215, B22214, B22213, B22212, B22211, B22210, B22209, B22208, B22207, B22206, B22205, B22204, B22203, B22202, B22201, B22200, B22199, B22198, B22197, B22196, B22195, B22194, B22193, B22192, B22191, B22190, B22189, B22188, B22187, B22186, B22185, B22184, B22172, B22171, B22170, B22169, B22168, B22167, B22166, B22165, B22164, B22163, B22162, B22161, B22160, B22159, B22158, B22157, B22156, B22120, B22119, B22118, B22117, B22116, B22115, B22114, B22113, B22112, B22111, B22110, B22109, B22108, B22107, B22106, B22105, B22104, B22103, B22102, B22101, B22100, B22099, B22098, B22097, B22096, B22095, B22094, B22093, B22092, B22091, B22090, B22089, B22088, B22087, B22086, B22085, B22084, B22083, B22082, B22081, B22080, B22079, B22078, B22077, B22076, B22075, B22074, B22073, B22072, B22071, B22070, B22069, B22068, B22067, B22066, B22065, B22064, B22063, B22062, B22061, B22060, B22059, B22058, B22057, B22056, B22055, B22054, B22053, B22052, B22051, B22050, B22049, B22048, B22047, B22046, B22045, B22044, B22043, B22042, B22041, B22040, B22039, B22038, B22037, B22036, B22035, B22034, B22033, B22032, B22031, B22030, B22029, B22028, B22027, B22026, B22025, B22024, B22023, B22022, B22021, B22020, B22019, B22018, B22017, B22016, B22015, B22014, B22013, B22012, B21975, B21974, B21973, B21972, B21971, B21970, B21969, B21968, B21967, B21966, B21965, B21964, B21963, B21962, B21961, B21960, B21959, B21958, B21898, B21897, B21896, B21895, B21894, B21893, B21892, B21891, B21890, B21810, B21809, B21808, B21807, B21806, B21805, B21804, B21803, B21787, B21786, B21785, B21784, B21783, B21782, B21781, B21780, B21779, B21744, B21743, B21742, B21741, B21740, B21739, B21738, B21737, B21736, B21735, B21734, B21733, B21732, B21731, B21730, B21729, B21728, B21727, B21726, B21725, B21724, B21723, B21722, B21721, B21720, B21719, B21718, B21717, B21716, B21715, B21714, B21713, B21712, B21711, B21710, B21709, B21708, B21707, B21706, B21705, B21704, B21703, B21702, B21701, B21700, B21699, B21698, B21697, B21696, B21695, B21694, B21634, B21633, B21632, B21631, B21630, B21629, B21628, B21627, B21626, B21625, B21624, B21623, B21622, B21621, B21620, B21619, B21618, B21595, B21594, B21593, B21592, B21591, B21590, B21589, B21588, B21587, B21586, B21585, B21584, B21583, B21582, B21581, B21580, B21579, B21578, B21577, B21576, B21575, B21574, B21549, B21514, B21513, B21512, B21501, B21500, B21499, B21498, B21497, B21496, B21495, B21494, B21493, B21473, B21472, B21471, B21470, B21469, B21468, B21467, B21466, B21465, B21464, B21463, B21385, B21384, B21383, B21382, B21381, B21380, B21379, B21378, B21348, B21347, B21346, B21345, B21344, B21343, B21342, B21341, B21340, B21339, B21338, B21337, B21336, B21335, B21334, B21316, B21315, B21314, B21313, B21312, B21302, B21301, B21300, B21299, B21298, B21297, B21296, B21295, B21294, B21293, B21292, B21291, B21290, B21207, B21206, B21205, B21204, B21203, B21202, B21201, B21199, B21198, B21197, B21196, B21195, B21189, B21188, B21187, B21186, B21185, B21184, B21183, B21182, B21181, B21180, B21143, B21142, B21141, B21140, B21139, B21138, B21137, B21136, B21135, B21134, B21133, B21132, B21131, B21075, B21074, B21073, B21072, B21071, B21070, B20969, B20968, B20967, B20966, B20965, B20964, B20963, B20962, B20961, B20960, B20959, B20958, B20957, B20956, B20955, B20954, B20943, B20942, B20941, B20940, B20939, B20938, B20937, B20936, B20935, B20934, B20933, B20932, B20931, B20930, B20929, B20928, B20927, B20926, B20925, B20924, B20923, B20922, B20921, B20920, B20919, B20918, B20917, B20916, B20915, B20914, B20913, B20912, B20911, B20910, B20909, B20908, B20907, B20906, B20905, B20904, B20903, B20902, B20901, B20900, B20899, B20898, B20897, B20896, B20895, B20894, B20893, B20892, B20891, B20890, B20889, B20866, B20865, B20764, B20763, B20762, B20761, B20760, B20759, B20758, B20757, B20756, B20755, B20754, B20753, B20752, B20751, B20750, B20749, B20748, B20747, B20746, B20745, B20744, B20676, B20675, B20674, B20673, B20672, B20671, B20670, B20669, B20589, B20588, B20587, B20586, B20585, B20584, B20583, B20582, B20581, B20580, B20579, B20578, B20577, B20576, B20575, B20513, B20512, B20511, B20465, B20444, B20443, B20442, B20441, B20440, B20439, B20438, B20437, B20436, B20435, B20434, B20266, B20265, B20264, B20263, B20184, B20183, B20182, B20181, B20180, B20179, B20178, B20177, B20176, B20175, B20174, B20173, B20172, B20171, B20170, B20169, B20168, B20167, B20166, B20165, B20002, B20001, B20000, B19999, B19998, B19997, B19996, B19995, B19994, B19993, B19992, B19991, B19591, B19590, B19589, B19588, B19587, B19586, B19585, B19584, B19583, B19431, B19430, B19429, B19428, B19427, B19426, B19425, B19424, B19302, B19301, B19300, B19299, B19298, B19297, B19296, B19295, B19294, B19293, B19292, B19291, B19290, B19289, B19288, B19287, B19286, B19285, B19284, B19283, B19282, B19281, B19280, B19279, B19278, B19277, B19276, B19275, B19274, B19273, B19272, B19271, B19270, B19269, B19268, B19267, B19266, B19265, B19264, B19263, B19262, B19261, B19260, B19259, B19258, B19257, B19256, B19255, B19254, B19253, B19252, B19251, B19250, B19249, B19248, B19247, B19246, B19245, B19244, B19243, B19242, B19241, B19240, B19239, B19238, B19237, B19236, B19235, B19234, B19233, B19232, B19231, B19230, B19125, B19124, B19123, B19122, B19121, B19120, B19119, B19118, B19117, B19116, B19115, B19114, B19113, B19112, B19111, B19110, B19109, B19108, B19107, B19106, B19105, B19104, B19103, B19102, B19101, B19100, B19099, B19098, B19097, B19096, B19066, B19065, B19064, B19063, B19062, B19061, B19060, B19059, B19058, B19057, B19056, B19055, B19054, B19053, B19052, B19051, B19050, B19049, B19048, B19047, B19046, B19045, B19044, B19043, B19042, B19041, B19040, B19039, B19038, B19037, B19036, B18984, B18983, B18982, B18981, B18980, B18978, B18977, B18976, B18975, B18974, B18952, B18951, B18949, B18948, B18947, B18941, B18884, B18883, B18882, B18881, B18880, B18879, B18878, B18877, B18876, B18875, B18874, B18873, B18872, B18835, B18834, B18833, B18832, B18831, B18830, B18829, B18828, B18827, B18826, B18825, B18824, B18823, B18822, B18821, B18820, B18819, B18706, B18705, B18704, B18703, B18702, B18701, B18700, B18699, B18698, B18697, B18696, B18695, B18694, B18693, B18692, B18691, B18690, B18689, B18688, B18687, B18686, B18685, B18684, B18683, B18670, B18669, B18668, B18667, B18666, B18575, B18574, B18573, B18561, B18560, B18488, B18487, B18486, B18485, B18484, B18483, B18482, B18481, B18480, B18479, B18478, B18477, B18476, B18475, B18474, B18331, B18205, B18204, B18203, B18202, B17830, B17829, B17824, B17759, B17739, B17738, B17737, B17736, B17735, B17734, B17683, B17682, B17681, B17680, B17679, B17678, B17677, B17676, B17675, B17616, B17615, B17614, B17613, B17612, B17611, B17610, B17461, B17334, B17333, B17332, B17331, B17330, B17295, B17294, B17293, B17292, B17179, B17178, B17177, B16999, B16998, B16997, B16996, B16995, B16994, B16934, B16933, B16932, B16931, B16731, B16730, B16729, B16728, B16727, B16726, B16725, B16724, B16723, B16702, B16701, B16700, B16699, B16698, B16697, B16696, B16695, B16694, B16693, B16692, B16691, B16690, B16689, B16682, B16641, B16640, B16639, B16638, B16637, B16341, B16340, B16339, B16338, B16015, B16014, B16013, B16012, B16011, and B16010

Distribution

Worldwide distribution:USA (nationwide) and countries including: Argentina, Bertec, Brasil, Brazil, Canada, Chile, China, India, Latin America, Mexico,Osteonics SA, Polska, Portugal, Puerto Rico, and Singapore.

Quantity

999