FDA Enforcement Class II Terminated

NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-304 Lactate 45, Potassium 2, Sodium 140, Calcium 3,Magnesium 1, Chloride 101, Glucose (mg/dL) 100 Batch Size (liters) 60 The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch.

Recall: Z-1857-2014 · Reported July 2, 2014

Enforcement

Recall Number
Z-1857-2014
Event ID
68354
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
NxStage Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 2, 2014
Initiation Date
May 14, 2014
Classification Date
June 24, 2014
Termination Date
March 13, 2017
Address
350 Merrimack Street, N/A, Lawrence, MA, 01843-1748, United States

Description

NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-304 Lactate 45, Potassium 2, Sodium 140, Calcium 3,Magnesium 1, Chloride 101, Glucose (mg/dL) 100 Batch Size (liters) 60 The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch.

Reason

SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification

Code Info

Lot code: 40179091 40179069 40179068 40179029 40179019 31279094 3067901 3057932 3057919 3057918 3057909 3057908

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to France, Netherlands, Spain, Sweden, and the UK.

Quantity

16,625 cases