FDA Enforcement Class II Terminated

NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-302 Lactate , 40, Potassium 1, Sodium 140, Calcium 3, Magnesium 1, Chloride 105, Glucose (mg/dL) 100, Batch Size (liters): 60 The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch.

Recall: Z-1855-2014 · Reported July 2, 2014

Enforcement

Recall Number
Z-1855-2014
Event ID
68354
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
NxStage Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 2, 2014
Initiation Date
May 14, 2014
Classification Date
June 24, 2014
Termination Date
March 13, 2017
Address
350 Merrimack Street, N/A, Lawrence, MA, 01843-1748, United States

Description

NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-302 Lactate , 40, Potassium 1, Sodium 140, Calcium 3, Magnesium 1, Chloride 105, Glucose (mg/dL) 100, Batch Size (liters): 60 The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch.

Reason

SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification

Code Info

Lot codes: 40179127 40179107 40179109 40179105 40179093 40179034 31279120 31279118 3057921 3057920 3057915 3057914 3057906 3057905 3047924 3047923

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to France, Netherlands, Spain, Sweden, and the UK.

Quantity

19423 cases