Esthetic Abutment CC RP 1.5 mm pre-manufactured prosthetic components directly connected to the endosseous implant, catalog no. 36669 - Product Usage: Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function.
Enforcement
- Recall Number
- Z-1854-2020
- Event ID
- 85159
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Nobel Biocare Usa Llc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- May 13, 2020
- Initiation Date
- August 9, 2019
- Classification Date
- May 4, 2020
- Termination Date
- December 1, 2021
- Address
- 22715 Savi Ranch Pkwy, N/A, Yorba Linda, CA, 92887-4609, United States
Description
Esthetic Abutment CC RP 1.5 mm pre-manufactured prosthetic components directly connected to the endosseous implant, catalog no. 36669 - Product Usage: Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function.
Packages of abutments may contain an incorrect screw. While the screw will tighten, it cannot be guaranteed that the screw will not loosen. In case the screw loosens the abutment should be retightened with the correct screw.
Catalog no. 36669, Lot no. 13077560, UDI: (01)07332747032598(10)13077560
Worldwide distribution - US Nationwide distribution including in the states of AZ, AR, CA, CO, FL, GA, IL, IN, ID, KS, LA, MI, MD, ME, MA, MO, NY, NC, NJ, NE, NM, OH, OR, OK, PA, TX, TN, VA, WY, WA and WI. Countries of CA, CH, DE, ES, FR, GB, HU, IT, BR, CO, MX, CN, HK, IN, JP and TW.
227 pieces