FDA Enforcement Class II Terminated

Temporary Abutment Engaging CC RP pre-manufactured prosthetic components directly connected to the endosseous implant, catalog no. 36664 - - Product Usage: Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function.

Recall: Z-1853-2020 · Reported May 13, 2020

Enforcement

Recall Number
Z-1853-2020
Event ID
85159
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Nobel Biocare Usa Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
May 13, 2020
Initiation Date
August 9, 2019
Classification Date
May 4, 2020
Termination Date
December 1, 2021
Address
22715 Savi Ranch Pkwy, N/A, Yorba Linda, CA, 92887-4609, United States

Description

Temporary Abutment Engaging CC RP pre-manufactured prosthetic components directly connected to the endosseous implant, catalog no. 36664 - - Product Usage: Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function.

Reason

Packages of abutments may contain an incorrect screw. While the screw will tighten, it cannot be guaranteed that the screw will not loosen. In case the screw loosens the abutment should be retightened with the correct screw.

Code Info

Catalog no. 36664, Lot no. 13077622, UDI: (01)07332747034301(10)13077622

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AZ, AR, CA, CO, FL, GA, IL, IN, ID, KS, LA, MI, MD, ME, MA, MO, NY, NC, NJ, NE, NM, OH, OR, OK, PA, TX, TN, VA, WY, WA and WI. Countries of CA, CH, DE, ES, FR, GB, HU, IT, BR, CO, MX, CN, HK, IN, JP and TW.

Quantity

239