FDA Enforcement Class II Ongoing

Leica Surgical Operating Microscopes, M530 OHX, Part Number 10448737

Recall: Z-1850-2022 · Reported October 5, 2022

Enforcement

Recall Number
Z-1850-2022
Event ID
90818
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Leica Microsystems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 5, 2022
Initiation Date
August 26, 2022
Classification Date
September 29, 2022
Address
10 Parkway North blvd ste 300, N/A, deerfield, IL, 60089-6622, United States

Description

Leica Surgical Operating Microscopes, M530 OHX, Part Number 10448737

Reason

During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.

Code Info

Serial Numbers: 270422001, 160322001, 230222001, 140422001, 100122002, 60721001, 290522001, 310522001, 81121001, 290921001, 130122001, 111021001, 131021001, 171021001, 10622002, 280222002, 51121001, 90821001, 170222001, 170522001, 11221001, 310322001, 260522001, 270522001, 140222001, 270422002, 30122001, 70122001, 271121001, 300322001, 31221001, 41221001, 230222002, 221121001, 70422001, 60122001, 120522001, 90921001, 200122001, 140122001, 100122001, 100721001, 100721002, 110721001, 20522001, 30921001, 261021001, 210422001, 40422001, 240122001, 140721001, 30821001, 50821001, 280721001, 290721001, 300721001, 300821001, 161021001, 160921002, 110222001, 91121001, 141221001, 151221001, 111221001, 301221001, 120222001, 120721001, 160122001, 291221001, 200721001, 91221001, 20921001, 220621001, 131121001, 161121001, 80422001, 240522001, 190821001, 300921001, 61021001, 80721001, 160921001, 10921001, 270122001, 90122001, 10622001, 50122001, 80422002, 211021001, 251021001, 200522001, 110522001, 130522001, 140522001, 110821002, 60522001, 200821001, 210921001, 20322002, 70322001, 100322002, 190322001, 30322001, 301021001, 30522001, 50721001, 261221001, 111121001, 81121002, 261121001, 301121001, 130821001, 60422002, 11221001, 140222002

Distribution

US Nationwide distribution.

Quantity

115 devices