FDA Enforcement
Class II
Ongoing
Leica Surgical Operating Microscopes, Arveo, Part Number 10448999
Recall: Z-1849-2022
·
Reported October 5, 2022
Enforcement
- Recall Number
- Z-1849-2022
- Event ID
- 90818
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Leica Microsystems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 5, 2022
- Initiation Date
- August 26, 2022
- Classification Date
- September 29, 2022
- Address
- 10 Parkway North blvd ste 300, N/A, deerfield, IL, 60089-6622, United States
Description
Leica Surgical Operating Microscopes, Arveo, Part Number 10448999
Reason
During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.
Code Info
Serial Numbers: 210721001, 180721001, 190721001, 121021001, 100821001, 60821001, 90821001, 120821001, 10821001, 80721001, 40821001, 130821001, 290821001, 240821001, 91021001, 180921001, 181121001, 151021001, 190821001, 160921001, 20721001, 50721001, 40921001, 150821001, 160821001
Distribution
US Nationwide distribution.
Quantity
25 devices