FDA Enforcement Class II Terminated

The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical purposes, for placing a cable incorporating an irradiated iridium seed internally or close by a malignant tumor or tumor bed in a practice known as brachytherapy.

Recall: Z-1849-2015 · Reported July 8, 2015

Enforcement

Recall Number
Z-1849-2015
Event ID
71451
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Varian Medical Systems Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
July 8, 2015
Initiation Date
May 18, 2015
Classification Date
June 26, 2015
Termination Date
September 12, 2016
Address
501 Locust Ave, Charlottesville, VA, 22902-4869, United States

Description

The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical purposes, for placing a cable incorporating an irradiated iridium seed internally or close by a malignant tumor or tumor bed in a practice known as brachytherapy.

Reason

After an application freeze and restart of the VariSource iX series by power cycling, the Partial Fraction generated by the system will not be correct. The application freeze issue affects only the systems equipped with magnetic Hard Disk Drives (HDD).

Code Info

JAQ

Distribution

Worldwide Distribution.

Quantity

55