FDA Enforcement Class II Ongoing

Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143

Recall: Z-1844-2024 · Reported May 29, 2024

Enforcement

Recall Number
Z-1844-2024
Event ID
94508
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips North America Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 29, 2024
Initiation Date
May 3, 2024
Classification Date
May 17, 2024
Address
222 Jacobs St, Cambridge, MA, 02141-2289, United States

Description

Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143

Reason

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Code Info

UDI-DI: N/A Serial Numbers: 42246 85713 42029 71774 85668 85680 85685 85694 85705 42023 42024 42375 42217 42402 71044 85698 85700 85146 85677 85699 71595 71815 85649 42320 85684 71836 85679 85682 85697 85701 85708 85714

Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.