EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table

Recall: Z-1844-2016 · Reported June 8, 2016

Enforcement

Recall Number
Z-1844-2016
Event ID
74128
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Skytron, Div. The KMW Group, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 8, 2016
Initiation Date
April 6, 2016
Classification Date
May 31, 2016
Termination Date
July 20, 2017
Address
5085 Corporate Exchange Blvd SE, N/A, Grand Rapids, MI, 49512-5515, United States

Description

EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table

Reason

The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.

Code Info

Model number 3502. All units. Units entered commercial distribution between April 30, 2014 and July 24, 2014

Distribution

Nationwide Distribution to AL, AZ, CO, IA, WI, PA, TN, MA, TX, MI, FL, MS, LA, and CA.

Quantity

458