EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table
Enforcement
- Recall Number
- Z-1844-2016
- Event ID
- 74128
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Skytron, Div. The KMW Group, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 8, 2016
- Initiation Date
- April 6, 2016
- Classification Date
- May 31, 2016
- Termination Date
- July 20, 2017
- Address
- 5085 Corporate Exchange Blvd SE, N/A, Grand Rapids, MI, 49512-5515, United States
Description
EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table
The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.
Model number 3502. All units. Units entered commercial distribution between April 30, 2014 and July 24, 2014
Nationwide Distribution to AL, AZ, CO, IA, WI, PA, TN, MA, TX, MI, FL, MS, LA, and CA.