Hip Preservation System Signature Series PreBent Burs under the following labels: 1) PreBent Spherical Bur, 5.5mm x 19cm Catalog Number HPS-HB11, 2) PreBent Polishing Bur, 5.5mm x 19cm Catalog Number HPS-HB12 and 3) PreBent Oval Bur, 6.0mm x 19cm Catalog Number HPS-HB13 The ConMed Hip Preservation System Signature Series is intended to be used for resection of osseous tissue (bone) in arthroscopic and endoscopic surgical procedures
Enforcement
- Recall Number
- Z-1842-2016
- Event ID
- 73123
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- ConMed Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 8, 2016
- Initiation Date
- December 21, 2015
- Classification Date
- May 31, 2016
- Termination Date
- May 24, 2018
- Address
- 525 French Rd, N/A, Utica, NY, 13502-5945, United States
Description
Hip Preservation System Signature Series PreBent Burs under the following labels: 1) PreBent Spherical Bur, 5.5mm x 19cm Catalog Number HPS-HB11, 2) PreBent Polishing Bur, 5.5mm x 19cm Catalog Number HPS-HB12 and 3) PreBent Oval Bur, 6.0mm x 19cm Catalog Number HPS-HB13 The ConMed Hip Preservation System Signature Series is intended to be used for resection of osseous tissue (bone) in arthroscopic and endoscopic surgical procedures
Metal shavings released from burs during use are due to contact between the bur and the hood of the PreBent Bur assembly. During the manufacturing process an assembly error occurred which caused inadvertent mixing of two sizes of bearings.
Lot Numbers: 1) 569449, 577583, 577969, 579655, 580365, 581176, 581198, 582216, 587320, 594569, 658219, 669615, 673913 & 688136; 2) 569462, 577800, 577971 & 579652; 3) 569468, 577835, 577974, 579650, 583736, 583742, 587514, 627923, 659317 & 664751.
Worldwide Distribution - Nationwide to CA, CO, FL, IN, MD, MN, NC, NY, OH, TX, UT & VA; International: Canada, Wales
991 pieces