FDA Enforcement Class II Terminated

RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 .

Recall: Z-1838-2021 · Reported June 16, 2021

Enforcement

Recall Number
Z-1838-2021
Event ID
87977
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
RAYSEARCH LABORATORIES AB
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 16, 2021
Initiation Date
April 21, 2021
Classification Date
June 9, 2021
Termination Date
July 27, 2023
Address
Sveavagen 9, N/A, Stockholm, N/A, N/A, Sweden

Description

RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 .

Reason

Evaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. It also describes an issue with contouring when the spacing between slices is d 1 mm, present in RayStation/RayPlan versions from 6 up to 10A.

Code Info

Software version: RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B, RayPlan 2 including some service packs.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AZ, CA, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Canada, Colombia, Czech Republic, Denmark, Egypt, Estonia, France, French Polynesia, Germany, Hong Kong, China, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zealand, Norway, Poland, Qatar, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, UK,

Quantity

Domestic: 13,722 systems