FDA Enforcement Class II Terminated

PROBEAT WITH DISCRETE SPOT SCANNING SYSTEM Product Usage: Hitachi PROBEAT with DSSS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Recall: Z-1838-2015 · Reported July 29, 2015

Enforcement

Recall Number
Z-1838-2015
Event ID
71056
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hitachi America, Ltd., Power Systems Division
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 29, 2015
Initiation Date
April 21, 2015
Classification Date
July 17, 2015
Termination Date
March 24, 2016
Address
1840 Old Spanish Trl, N/A, Houston, TX, 77054-2002, United States

Description

PROBEAT WITH DISCRETE SPOT SCANNING SYSTEM Product Usage: Hitachi PROBEAT with DSSS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Reason

At a PROBEAT III site installed in Japan, the operator of the system found that the radiation was delivered inconsistent with the treatment plan and stopped the irradiation. The irradiation control system may be potentially affected when rebooting a part of the control system.

Code Info

No serial number or lot code. Each system is designed for a specific facility therefore no codes are needed for traceability

Distribution

Distribution in US (Texas) and Japan

Quantity

3