The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It contains signaling electronics to indicate the presence of the grid and detector as well as detector orientation. It may be mounted to a wall stand or incorporated in a radiographic table. It is intended to be used in general radiographic applications as a component of a radiographic system.
Enforcement
- Recall Number
- Z-1836-2014
- Event ID
- 67715
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Claymount Assembli
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 9, 2014
- Initiation Date
- March 15, 2014
- Classification Date
- June 30, 2014
- Termination Date
- October 18, 2018
- Address
- Anholtseweg 44, Dinxperlo, N/A, Netherlands
Description
The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It contains signaling electronics to indicate the presence of the grid and detector as well as detector orientation. It may be mounted to a wall stand or incorporated in a radiographic table. It is intended to be used in general radiographic applications as a component of a radiographic system.
Claymount Assemblies BV is recalling the Harmony2 model of Harmony Bucky because 1) the devices are incorrectly labeled in that the label does not list the manufacturer address, the date of manufacture, or a statement of compliance with 21 CFR Subchapter J, and 2) the device's insulation values do not comply with IEC 60601-1 in all cases.
*** US Devices Only *** 1) Part ID: 17360; Serial Numbers: WO186722/E-1 , WO196470/E-1 , WO198534/E-1. 2) Part ID: 17361; Serial Numbers: WO186723/E-1. 3) Part ID: 17362; Serial Numbers: WO186724/E-1 , WO196360/E-1 , WO198545/E-2. 4) Part ID: 18476; Serial Numbers: WO196188/E-1 , WO196188/E-2 , WO196189/E-2. 5) Part ID: 18477; Serial Numbers: WO198547/E-1. 6) Part ID: 18479; Serial Numbers: WO198552/E-1.
Worldwide Distribution - USA including NC, OH, CO, CA, NJ, IL, OH, PA, FL and Internationally to Canada, China, France, Italy, Netherlands, South Africa, Switzerland, Turkey, and United Kingdom.
12 devices