FDA Enforcement
Class II
Terminated
NIPRO Safe Touch Safety Scalp Vein Set
Recall: Z-1834-2014
·
Reported June 25, 2014
Enforcement
- Recall Number
- Z-1834-2014
- Event ID
- 65274
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Nipro Medical Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 25, 2014
- Initiation Date
- May 4, 2013
- Classification Date
- June 19, 2014
- Termination Date
- March 2, 2015
- Address
- 3150 Nw 107th Ave, N/A, Miami, FL, 33172, United States
Description
NIPRO Safe Touch Safety Scalp Vein Set
Reason
There is a possibility of a crack in the CP luer connector. The position of the crack is at the connection site of the CP luer connector and the tubing. The crack could cause or contribute to leakage of fluids.
Code Info
Product Code: PR+25G19, Lots #s 12H31H and 12I14H; Product Code: SPR+21G19, Lots #s 12I24 and 12J11; Product Code: PR-25G19, Lot # 12H27H; and PR-23G19, Lot # 12I12H.
Distribution
Worldwide Distribution - USA and the countries of Canada, Bolivia, Chile, Uruguay, Belize, Dominica, Honduras, Jamaica, Paraguay and Venezuela.
Quantity
176,000 pieces