FDA Enforcement Class II Terminated

The Plum 360 is a large volume infuser capable of delivering fluids for a variety of therapies such as parenteral, enteral, or epidural infusions. The Plum 360 infuser can deliver fluids over a broad range of infusion rates and is capable of Concurrent delivery from one or more rigid or flexible fluid containers. The Plum 360 infuser features Concurrent, Secondary, and Piggyback infusions. A positive valving cassette allows two lines to be delivered at independent rates. The volume to be infused (VTBI) is delivered through one line to a patient. The two lines can be delivered in Concurrent mode (together) or Piggyback mode (one after another) without raising or lowering I.V. bags. The Plum 360 infuser also enables fluid pathway troubleshooting such as removing proximal air in line, without is disconnecting the patient line. The Plum 360 can act as a stand-alone infuser, or in conjunction with the Hospira MedNet software to provide medication safety software at the point of care, with customized drug libraries to support hospital defined protocols by clinical care area. In such a configuration, the Plum 360 infuser can communicate with systems on the network via Ethernet or state of the art wireless communication using an 802.11 a/b/g/n/, 2.4 GHz/5 GHz dual-band radio. The Plum 360 infuser and Hospira MedNet software interface with other hospital systems such as Electronic Health records, Electronic Medication Administration Records, Bar Code Point of Care, Real Time Location Services, and there systems. Each infuser includes a Connectivity Engine (CE) which provides both wired Ethernet and wireless 802.11 a/b/g/ networking capabilities. The Plum 360 infuser interfaces with Hospira MedNet application software to download drug library and infuser software updates and enable auto-programming of the infuser.The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Recall: Z-1832-2016 · Reported June 15, 2016

Enforcement

Recall Number
Z-1832-2016
Event ID
73799
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hospira Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 15, 2016
Initiation Date
March 25, 2016
Classification Date
June 3, 2016
Termination Date
March 6, 2018
Address
275 N Field Dr, N/A, Lake Forest, IL, 60045-2579, United States

Description

The Plum 360 is a large volume infuser capable of delivering fluids for a variety of therapies such as parenteral, enteral, or epidural infusions. The Plum 360 infuser can deliver fluids over a broad range of infusion rates and is capable of Concurrent delivery from one or more rigid or flexible fluid containers. The Plum 360 infuser features Concurrent, Secondary, and Piggyback infusions. A positive valving cassette allows two lines to be delivered at independent rates. The volume to be infused (VTBI) is delivered through one line to a patient. The two lines can be delivered in Concurrent mode (together) or Piggyback mode (one after another) without raising or lowering I.V. bags. The Plum 360 infuser also enables fluid pathway troubleshooting such as removing proximal air in line, without is disconnecting the patient line. The Plum 360 can act as a stand-alone infuser, or in conjunction with the Hospira MedNet software to provide medication safety software at the point of care, with customized drug libraries to support hospital defined protocols by clinical care area. In such a configuration, the Plum 360 infuser can communicate with systems on the network via Ethernet or state of the art wireless communication using an 802.11 a/b/g/n/, 2.4 GHz/5 GHz dual-band radio. The Plum 360 infuser and Hospira MedNet software interface with other hospital systems such as Electronic Health records, Electronic Medication Administration Records, Bar Code Point of Care, Real Time Location Services, and there systems. Each infuser includes a Connectivity Engine (CE) which provides both wired Ethernet and wireless 802.11 a/b/g/ networking capabilities. The Plum 360 infuser interfaces with Hospira MedNet application software to download drug library and infuser software updates and enable auto-programming of the infuser.The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Reason

Firm's investigation determined that the Proximal Occlusion Alarms observed in the field were attributed to the mechanism calibration process being performed with an out of specification metal cassette, resulting in an offset in proximal pressure.

Code Info

PLum 360 devices with the following list number: 300100405, 300100406, 300100605 , 300101305 , 300102705 , 300105405 and 300106507.

Distribution

U.S. distribution : Natrionwide. Foreign :Brazil, Canada, Chile, Colombia, Ecuador, Guatemala, Mexico, Oman, Peru, Taiwan, Turkey, United Kingdom and Uruguay.

Quantity

1490 infusion pumps.