FDA Enforcement Class II Terminated

Normand Remisol Advance Data Manager is a software device that interfaces between laboratory information systems (LIS) and laboratory instruments.

Recall: Z-1831-2021 · Reported June 16, 2021

Enforcement

Recall Number
Z-1831-2021
Event ID
87750
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Normand-Info S.A.S.U.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 16, 2021
Initiation Date
March 22, 2021
Classification Date
June 8, 2021
Termination Date
December 30, 2022
Address
7 rue Frederic Degeorge, N/A, Arras CEDEX, N/A, N/A, France

Description

Normand Remisol Advance Data Manager is a software device that interfaces between laboratory information systems (LIS) and laboratory instruments.

Reason

Instructions For Use for software versions 1.5 to 1.9 did not provide sufficient instructions regarding the proper programming and testing of rules. Therefore, rules written by users that were not validated or incorrectly validated may lead to erroneous results.

Code Info

Catalog Numbers: B08688; B22517; B35058; B74811; B92487 UDI: None Lot Numbers: All Lots

Distribution

Worldwide distribution - US Nationwide distribution in the states of Alabama, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Indiana, Iowa, Kansas, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, Pennsylvania, Rhode Island, South Carolina, Texas, West Virginia and Wyoming. The countries of Australia, Austria, Belgium, Brunei, Canada, China, Colombia, Congo, Czech Republic, Denmark, Ethiopia, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Lithuania, Malaysia, Mayotte, Morocco, Netherlands, Netherlands Antillean, New Zealand, Norway, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, UAE, UK, and Zimbabwe.

Quantity

8,402 units with the affected software