FDA Enforcement Class II Ongoing

Ingenia 3.0T with Evolution Upgrade 3.0T - - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782117

Recall: Z-1830-2024 · Reported May 22, 2024

Enforcement

Recall Number
Z-1830-2024
Event ID
94493
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips North America Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 22, 2024
Initiation Date
April 22, 2024
Classification Date
May 15, 2024
Address
222 Jacobs St, N/A, Cambridge, MA, 02141-2289, United States

Description

Ingenia 3.0T with Evolution Upgrade 3.0T - - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782117

Reason

Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation

Code Info

UDI-DI: (01)00884838099739(21) Serial Numbers: 42029 42217 42246 71044 71774 85146 85645 85663 85664 85668 85677 85680 85681 85685 85700 85704 85705 85706 85713

Distribution

Worldwide - US Nationwide distribution in the states of CA, KS, MS, NH and the countries of Austria, Croatia, Finland, France, Germany, Italy, Lebanon, Netherlands, Norway, Sweden, Switzerland, Thailand, United Kingdom.

Quantity

19 units