FDA Enforcement Class II Terminated

Custom Cardiovascular Procedure Kits; Over-Pressure Safety Valves. For use only in the extracorporeal circuit for the cardiopulmonary bypass procedures for which the user designed it.

Recall: Z-1828-2015 · Reported July 1, 2015

Enforcement

Recall Number
Z-1828-2015
Event ID
71196
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Terumo Cardiovascular Systems Corp
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
July 1, 2015
Initiation Date
May 6, 2015
Classification Date
June 22, 2015
Termination Date
July 27, 2015
Address
28 Howe St, Ashland, MA, 01721-1305, United States

Description

Custom Cardiovascular Procedure Kits; Over-Pressure Safety Valves. For use only in the extracorporeal circuit for the cardiopulmonary bypass procedures for which the user designed it.

Reason

The Overpressure Safety Valve is not cleared with an indication for use on the arterial side of the extracoporeal bypass circuit. It is intended for use only on the venous side. There is a remote possibility of minimal blood loss, fluid weeping, or dripping on the user.

Code Info

Part Number: 66086-01: Lot Numbers: RG09, RM20 and Part Number: 70700-03: Lot Numbers: QL09, RH07, RL15, QG17, RD03

Distribution

Distributed to the states of MO and FL.

Quantity

163 devices