FDA Enforcement
Class II
Terminated
Leksell Vantage Stereotactic System. Instrument Carrier, Part of Leksell Vantage Arc (1053958). Neurological stereotaxic Instrument
Recall: Z-1826-2020
·
Reported May 6, 2020
Enforcement
- Recall Number
- Z-1826-2020
- Event ID
- 85352
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Elekta Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 6, 2020
- Initiation Date
- March 30, 2020
- Classification Date
- April 29, 2020
- Termination Date
- April 10, 2024
- Address
- 1450 Beale St Ste 205, N/A, Saint Charles, MO, 63303, United States
Description
Leksell Vantage Stereotactic System. Instrument Carrier, Part of Leksell Vantage Arc (1053958). Neurological stereotaxic Instrument
Reason
A faulty locking piece of the Instrument Carrier that does not fulfil the requirement of locking force to the Arc has been found during an internal check of a Leksell Vantage Stereotactic Arc System.
Code Info
Serial Numbers:SH00106, SH00120, SH00149, SH00160, SH00165, SH00110, SH00143, SH00150, SH00161, SH00166, SH00111, SH00144, SH00157, SH00162, SH00167, SH00112, SH00147, SH00158, SH00163, SH00168, SH00114, SH00148, SH00159, SH00164 and SH00178. UDI 7340048304887.
Distribution
US nationwide distributions in the states of MA, NH, NC, UT, OH, FL, MN, NY, LA and MT.
Quantity
25 units.