FDA Enforcement Class II Ongoing

Philips Spectral CT on Rails. Model Number: 728334.

Recall: Z-1825-2026 · Reported April 22, 2026

Enforcement

Recall Number
Z-1825-2026
Event ID
98588
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
PHILIPS MEDICAL SYSTEMS
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 22, 2026
Initiation Date
March 7, 2026
Classification Date
April 16, 2026
Address
222 Jacobs St, N/A, Cambridge, MA, 02141-2289, United States

Description

Philips Spectral CT on Rails. Model Number: 728334.

Reason

Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.

Code Info

Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.

Distribution

Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.

Quantity

3 systems