Philips Spectral CT on Rails. Model Number: 728334.
Enforcement
- Recall Number
- Z-1825-2026
- Event ID
- 98588
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- PHILIPS MEDICAL SYSTEMS
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 22, 2026
- Initiation Date
- March 7, 2026
- Classification Date
- April 16, 2026
- Address
- 222 Jacobs St, N/A, Cambridge, MA, 02141-2289, United States
Description
Philips Spectral CT on Rails. Model Number: 728334.
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.
Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.
3 systems