FDA Enforcement Class II Terminated

Tapered SwissPlus Implant System, Model No. SPB10

Recall: Z-1823-2017 · Reported April 26, 2017

Enforcement

Recall Number
Z-1823-2017
Event ID
76746
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer Dental Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 26, 2017
Initiation Date
March 20, 2017
Classification Date
April 14, 2017
Termination Date
November 1, 2017
Address
1900 Aston Ave, N/A, Carlsbad, CA, 92008-7308, United States

Description

Tapered SwissPlus Implant System, Model No. SPB10

Reason

It was manufactured with the minor diameter of the internal thread that is below specification.

Code Info

Lot #63498673; UDI: (01) 008890240177 (10) 63498673 (17) 200731

Distribution

Worldwide: ZA, FR, IT, IR

Quantity

188 units